18th International Congress of Metrology 2017
DOI: 10.1051/metrology/201703001
|View full text |Cite
|
Sign up to set email alerts
|

Measurement of temperature and gas concentration in a flame using a FTIR imager

Abstract: Abstract. Nowadays, determining the temperature of flames is a challenging measurement in industry. The EMPIR project 14IND04 EMPRESS, in its WP4, address these temperature measurements. The Infrared Lab at the Physics Department of Universidad Carlos III of Madrid (UC3M) has developed a technique for measuring the temperature of a standard flame, in collaboration with the Centro Español de Metrología (CEM). An Imaging Fourier Transform Spectrometer (FTIR) has been used to acquire the emitted radiation coming … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2019
2019
2021
2021

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 1 publication
0
2
0
Order By: Relevance
“…Calibration temperatures were 180 °C and 400 °C. Since the spectral emission of the flame is concentrated in a narrow band in comparison to the full spectral range of the instrument, these blackbody temperatures were sufficient to calibrate the instrument for flame temperatures up to 2400 °C [31], The calculated emissivity of these BBs is 0.981 and 0.985, respectively. The BBs temperature has been measured with a RT HEITRONICS TRTII working at 3.9 μm, traceable to the national standards.…”
Section: Uc3mmentioning
confidence: 99%
“…Calibration temperatures were 180 °C and 400 °C. Since the spectral emission of the flame is concentrated in a narrow band in comparison to the full spectral range of the instrument, these blackbody temperatures were sufficient to calibrate the instrument for flame temperatures up to 2400 °C [31], The calculated emissivity of these BBs is 0.981 and 0.985, respectively. The BBs temperature has been measured with a RT HEITRONICS TRTII working at 3.9 μm, traceable to the national standards.…”
Section: Uc3mmentioning
confidence: 99%
“…Two blackbodies were used with an aperture large enough (70 mm diameter) to cover the whole field of view of the sub window used. Calibration temperatures were 180 • C and 400 • C. Since the spectral emission of the flame is concentrated in a narrow band in comparison to the full spectral range of the instrument, these blackbody temperatures were sufficient to calibrate the instrument for flame temperatures up to 2500 • C [3,17].…”
Section: Experimental Setup and Instrumentationmentioning
confidence: 99%