Abstract:The purpose of this study is(was) to investigate the shielding ratio of 1 ㎜Pb and the off axis ratio outside the field edge at depth of 1 ㎝ from a phantom surface for 6 MV photon beam. A dose of 180 cGy was delivered to a depth of 10 ㎝ for a 10×10 ㎠ and 15×15 ㎠ field in the SAD technique. The off axis ratio was calculated by measuring the dose of optically stimulated luminescent nanoDot dosimeters(OSLnDs) positioned at 2, 4 and 6 ㎝ from the field edge, and the center axis of field. And the shielding ratio of 1… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.