2017
DOI: 10.1109/jerm.2017.2765638
|View full text |Cite
|
Sign up to set email alerts
|

Measurement of Deep Tissue Implanted Antenna Efficiency Using a Reverberation Chamber

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
7
0

Year Published

2017
2017
2024
2024

Publication Types

Select...
4
2
1

Relationship

3
4

Authors

Journals

citations
Cited by 13 publications
(7 citation statements)
references
References 29 publications
(38 reference statements)
0
7
0
Order By: Relevance
“…This relatively deep implantation depth of 75 mm in high water content tissue will provide a worst case measurement of into-body performance, ensuring successful operation in clinical applications. To reduce the effect of cable radiation [25] a small 0603 case style 2.3-2.7 GHz chip balun was placed at the input terminals of the PFMD and the arrangement was such that there was a significant volume of muscle mimicking liquid surrounding the feed cable to dampen spurious cable radiation [26].…”
Section: Background and Cp Measurementsmentioning
confidence: 99%
“…This relatively deep implantation depth of 75 mm in high water content tissue will provide a worst case measurement of into-body performance, ensuring successful operation in clinical applications. To reduce the effect of cable radiation [25] a small 0603 case style 2.3-2.7 GHz chip balun was placed at the input terminals of the PFMD and the arrangement was such that there was a significant volume of muscle mimicking liquid surrounding the feed cable to dampen spurious cable radiation [26].…”
Section: Background and Cp Measurementsmentioning
confidence: 99%
“…1. The long cylindrical form factor of the layered phantom ensures that the dominating wave propagation from the antenna-phantom system is radial and antenna feed cable radiation effects are minimized [7].…”
Section: A Tissue Mimicking Phantomsmentioning
confidence: 99%
“…The two cylinder approach allows for liquid muscle and fat tissue mimicking regions and the skin layer helps investigate a more realistic scenario where there is a high loss dielectric boundary between fat and air. The muscle equivalent material is described in [7] and the fat equivalent liquid used was Diethylene glycol dibutyl ether [8], with both a good match for the parameters in [9] at 2.38 GHz (Table I).…”
Section: A Tissue Mimicking Phantomsmentioning
confidence: 99%
“…1) Implant antenna total radiation efficiency measurements in a reverberation chamber [21]. 2) Implant antenna far-field |S 21 | measurements.…”
Section: Simulation and Measurement Setupmentioning
confidence: 99%
“…The efficiency measurements were carried out in a reverberation chamber manufactured by Bluetest.se with a Rohde and Schwarz ZVB-8 Vector Network Analyser (VNA) [21]. The efficiency of each AUT within a model of the phantom testbed was also determined by simulation for both tissue liquid types using Sim4Life.…”
Section: B Implant Antenna Tre Measurements In a Reverberation Chambermentioning
confidence: 99%