Abstract:In order to guarantee the liability and accuracy of ultrasonic flaw detectors, a system is developed for measuring the characteristics of ultrasonic flaw detectors. The parameters which evaluate the characteristics of the pulse excitation circuit include pulse amplitude voltage, pulse rise time, pulse duration, pulse repetition frequency, peak frequency, bandwidth and output impedance. The parameters which evaluate the characteristics of the amplifier circuit include cut-off frequency, pass band, dynamic range… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.