Measurement of activity distribution in an Am-241 disc source using peeled-off Gafchromic EBT3 films

Abstract: Commercial alpha-emitting sources are fabricated mainly in a disc type. An alpha particle irradiator in Radiation Bioengineering Laboratory at Seoul National University was installed with an Am-241 disc source. Commercial Am-241 disc sources are fabricated by incorporating the radioactive element into a thin substrate layer. Those disc sources are utilized assuming that the radioactive element is uniformly distributed in the active layer of disc sources. In this study, we employed peeled-off Gafchromic EBT3 fi… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 10 publications
(4 citation statements)
references
References 25 publications
0
4
0
Order By: Relevance
“…The FWHM of the E&Z 210 Po source was measured to be 25.58 keV using alpha spectroscopy (ORTEC Alpha Aria). The radius of the source was measured using unlaminated radiochromic films (EBT3 Ashland, NJ) and was determined to be 1.60 ± 0.30 mm [24]. Following the alpha spectroscopy and unlaminated EBT3 measurements, the source substrate was mounted on the source subassembly for ionization current measurements.…”
Section: Po Sourcementioning
confidence: 99%
“…The FWHM of the E&Z 210 Po source was measured to be 25.58 keV using alpha spectroscopy (ORTEC Alpha Aria). The radius of the source was measured using unlaminated radiochromic films (EBT3 Ashland, NJ) and was determined to be 1.60 ± 0.30 mm [24]. Following the alpha spectroscopy and unlaminated EBT3 measurements, the source substrate was mounted on the source subassembly for ionization current measurements.…”
Section: Po Sourcementioning
confidence: 99%
“…The resultant photomicrographs were analyzed using ImagJ software and compared the performance of the two scanning methods. Since the available microscope in our laboratory has a camera with a 2 8 -pixel color depth, the image of the scanner has been reduced from 2 16 bit to 2 8 bit to facilitate the comparison with the optical microscope.…”
Section: ( ) ( ) = +mentioning
confidence: 99%
“…Polymer-based solid state nuclear track detectors, including the CR-39 detector and polycarbonate (Makrofol and Lexan) detectors, are being successfully applied in heavy ion measurements and neutron radiography [1][2][3][4][5]. On the other hand, self-developing radiochromic films such as Gafchrmoic EBT3 and HD-V2 are utilized effectively in energetic heavy ions dose mapping, which is a prerequisite for radiotherapy, in the range of 0.2 Gy to 10 Gy, where the radiochromic films are readout using a variety of film scanners [6][7][8].…”
Section: Introductionmentioning
confidence: 99%
“…Refs. [8][9][10][11][12][13][14][15][16]) had shown that the distribution of alpha particle tracks was neither uniform nor random, whatever the precaution made to make such a fact obvious. The reason may be any process from the decay to the registration on the detector surface.…”
Section: Introductionmentioning
confidence: 99%