Mean glandular dose estimation using MCNPX for a digital breast tomosynthesis system with tungsten/aluminum and tungsten/ x‐ray anode‐filter combinations

Abstract: Breast cancer screening with x-ray mammography, using one or two projection images of the breast, is an indispensible tool in the early detection of breast cancer in women. Digital breast tomosynthesis (DBT) is a 3D imaging technique that promises higher sensitivity and specificity in breast cancer screening at a similar radiation dose to conventional two-view screening mammography. In DBT a 3D volume is reconstructed with anisotropic voxels from a limited number of x-ray projection images acquired over a limi… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

4
27
1

Year Published

2009
2009
2023
2023

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 28 publications
(32 citation statements)
references
References 26 publications
(40 reference statements)
4
27
1
Order By: Relevance
“…From the results we can see that for both MLO and CC there is an increase in MGD when glandular ratio is higher, which is in agreement with other published data [8]. The results in Figure 5 confirm the linearity of MGD over glandular ratio described in literature [4], [10]. Linearity is valid for all simulated energies as presented in Tables 1 and 2.…”
Section: Simulation Setupsupporting
confidence: 81%
See 1 more Smart Citation
“…From the results we can see that for both MLO and CC there is an increase in MGD when glandular ratio is higher, which is in agreement with other published data [8]. The results in Figure 5 confirm the linearity of MGD over glandular ratio described in literature [4], [10]. Linearity is valid for all simulated energies as presented in Tables 1 and 2.…”
Section: Simulation Setupsupporting
confidence: 81%
“…Figures 3 and 4. Resulting values for MGD are lower than those previously reported [8], [9], [10]. The reasons for this are approximations and assumptions used in this work considering modeling of the phantom and source spectrum.…”
Section: Simulation Setupcontrasting
confidence: 46%
“…The variation in the normalized glandular dose coefficient with x-ray tube angle relative to the central projection (0°), referred to as relative glandular dose, was weakly dependent on fibroglandular fraction and x-ray spectrum and showed a larger dependence on breast size and thickness in mediolateral oblique (MLO) view compared with craniocaudal (CC) view (71). Patient positioning, particularly for MLO view, was reported to cause a 5%-13% variation in MGD (72). While the initial study (71) was based on a DBT system using molybdenum or rhodium target xray tube, subsequent studies addressed tungsten target x-ray tubes (72,73).…”
Section: Radiation Dosementioning
confidence: 99%
“…Patient positioning, particularly for MLO view, was reported to cause a 5%-13% variation in MGD (72). While the initial study (71) was based on a DBT system using molybdenum or rhodium target xray tube, subsequent studies addressed tungsten target x-ray tubes (72,73). Currently, many DBT systems use tungsten target with aluminum, silver, or rhodium filter (Table 1) and operate at a slightly higher applied tube voltage (kilovolts peak) than FFDM, partly to reduce radiation dose to the breast.…”
Section: Radiation Dosementioning
confidence: 99%
“…In a few studies, models of the breast compressed for the medio-lateral oblique (MLO) view have been used, but the models were subjectively derived. 18,19,[29][30][31] Although these models have proved appropriate, the development of more sophisticated models that better represent not only the "average" breast shape but also attempt to include the span of clinically realistic breast shape variations would be useful in many of the areas of research listed above that use software breast models.…”
Section: Introductionmentioning
confidence: 99%