<b>Defibrillation Testing and Early Neurologic Outcome</b>

Abstract: SUMMARYDuring implantable cardioverter-defibrillator (ICD) implantation, ventricular defibrillation testing (DFT) is considered a standard procedure. This procedure often requires multiple ventricular fibrillation (VF) inductions. These repeated short episodes of circulatory arrest with global cerebral ischemia may cause neurological damage. In the present study, patients undergoing initial ICD implantation and limited induction of VF for defibrillation safety margin testing were evaluated for pre-and postoper… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

1
12
0
1

Year Published

2010
2010
2018
2018

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 13 publications
(14 citation statements)
references
References 24 publications
1
12
0
1
Order By: Relevance
“…However, the clinical relevance of this transient finding is unclear because DT does not appear to cause cognitive dysfunction 24-48 hours following ICD implantation [200,201]. Although an increase in biochemical markers of myocardial injury can be observed during ICD implantation or after spontaneous clinical shocks, true intraoperative myocardial infarction (MI) is rare, even when extensive DT is performed [202][203][204][205].…”
Section: Iia B-nrmentioning
confidence: 99%
“…However, the clinical relevance of this transient finding is unclear because DT does not appear to cause cognitive dysfunction 24-48 hours following ICD implantation [200,201]. Although an increase in biochemical markers of myocardial injury can be observed during ICD implantation or after spontaneous clinical shocks, true intraoperative myocardial infarction (MI) is rare, even when extensive DT is performed [202][203][204][205].…”
Section: Iia B-nrmentioning
confidence: 99%
“…[8][9][10][11] Induction of ventricular fibrillation (VF) during DFT can cause hemodynamic compromise, clinical deterioration, and life-threatening cerebrovascular accidents. [12][13][14][15] Thus, we conducted a retrospective study to assess the effect of DFT testing on long-term outcomes in patients with an ICD or cardiac resynchronization therapy defibrillator (CRT-D).…”
Section: Mplantable Cardioverter-defibrillators (Icds) Have Beenmentioning
confidence: 99%
“…[11][12][13] Murkin et al 11 assessed 14 elective ICD patients in whom 12Ϯ6 episodes of VF were intraoperatively induced. At postoperative follow-up (5 days), 10 of 14 patients met the criteria for a significant 20% deficit in cognitive function from baseline, in particular, short-term memory and attention.…”
Section: Editorial See Page 118 Clinical Perspective On P 177mentioning
confidence: 99%
“…At postoperative follow-up (5 days), 10 of 14 patients met the criteria for a significant 20% deficit in cognitive function from baseline, in particular, short-term memory and attention. More recently, Karaoguz et al 13 assessed 16 ICD patients with 6 standardized cognitive function tests and NSE levels. Tests were administered 24 to 48 hours before implantation, and NSE levels were tested at 2, 24, and 48 hours after surgery.…”
Section: Editorial See Page 118 Clinical Perspective On P 177mentioning