2013
DOI: 10.1016/j.brs.2013.03.010
|View full text |Cite
|
Sign up to set email alerts
|

Longitudinal Impedance Variability in Patients with Chronically Implanted DBS Devices

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

10
48
1

Year Published

2015
2015
2022
2022

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 58 publications
(59 citation statements)
references
References 19 publications
10
48
1
Order By: Relevance
“…To the best of our knowledge, our study is the first to reveal a continuous decrease of therapeutic impedance values from 1036U to 786U within 10 years of STN-DBS, and it corresponds well to previous findings of a continuous decline of impedance over time [8]. As the stimulation settings remained constant over time and therefore cannot explain our findings, progressive neurodegeneration may be hypothesized as a possible explanation, because an associated increase of mean diffusivity and isotropy [9] would directly affect brain conductivity and hence impedance measurements [7].…”
Section: Discussionsupporting
confidence: 90%
“…To the best of our knowledge, our study is the first to reveal a continuous decrease of therapeutic impedance values from 1036U to 786U within 10 years of STN-DBS, and it corresponds well to previous findings of a continuous decline of impedance over time [8]. As the stimulation settings remained constant over time and therefore cannot explain our findings, progressive neurodegeneration may be hypothesized as a possible explanation, because an associated increase of mean diffusivity and isotropy [9] would directly affect brain conductivity and hence impedance measurements [7].…”
Section: Discussionsupporting
confidence: 90%
“…2). Thickness and conductivity of the encapsulation layer around the electrode, bulk tissue conductivity, and electrical stimulation activity itself contribute significantly to electrode impedance [20][21][22]. The time course of the transient changes in impedance may be related to microglial activation and subsequent fusion into multinuclear giant cells or mild to moderate inflammation as a potential mechanism of action for peri-implant impedance change with resulting GFAP positive pathology findings [23,24] (Fig.…”
Section: Discussionmentioning
confidence: 99%
“…The chronology of impedance fluctuations after DBS electrode implantation has been investigated by several groups in different experimental and clinical situations [7,8,14,28,29,30]. Clinically relevant impedance increases occur in the first few weeks after implantation, possibly as a result of changes at the electrode-brain interface [31].…”
Section: Discussionmentioning
confidence: 99%
“…The safety and efficacy of CC DBS in newly operated PD patients has been established [12]. The main benefit of CC stimulation is thought to be in the initial phase after surgery (weeks to months) when impedances change most rapidly, primarily as the result of changes at the electrode-brain tissue interface [7,8]. This may shorten the time needed to reach optimal programming [13].…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation