2008
DOI: 10.1093/bja/aen174
|View full text |Cite
|
Sign up to set email alerts
|

LMA-Supreme TM —a new single-use LMA TM with gastric access: a report on its clinical efficacy

Abstract: Insertion success, glottic seal pressure and gastric access were similar in SLMA and PLMA.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

21
117
4
1

Year Published

2009
2009
2017
2017

Publication Types

Select...
7
2

Relationship

0
9

Authors

Journals

citations
Cited by 137 publications
(143 citation statements)
references
References 11 publications
(13 reference statements)
21
117
4
1
Order By: Relevance
“…We assume that relatively larger size SLMA might have been the cause of difficult insertion in those 3 cases. The range of insertion time (TOI) in our study was 11-15 seconds, which is supported by similar findings by Verghese C et al [7] , Cook TM et al [11] and Theiler LG et al [14] . We think that the intelligent design of the fixation tube (FT) along with the elliptical and anatomically shaped semi rigid airway tube of the SLMA facilitated rapid insertion.…”
Section: Discussionsupporting
confidence: 92%
See 1 more Smart Citation
“…We assume that relatively larger size SLMA might have been the cause of difficult insertion in those 3 cases. The range of insertion time (TOI) in our study was 11-15 seconds, which is supported by similar findings by Verghese C et al [7] , Cook TM et al [11] and Theiler LG et al [14] . We think that the intelligent design of the fixation tube (FT) along with the elliptical and anatomically shaped semi rigid airway tube of the SLMA facilitated rapid insertion.…”
Section: Discussionsupporting
confidence: 92%
“…The FT was found, in early pilot studies, to act as a Visual guide to 'correct' size selection, that is after inflation of the cuff to a pressure of 60 cm H 2 O, the FT should be 1.5-2 cm from the upper lip, if the FT is less than this distance the size chosen may be too small and if 3.0 cm from the upper lip the size chosen may be too large. [7] It has been used in adult patients and performed well. [3][4][5] The latest advent of LMA Supreme is its paediatric versions which are now available in four different sizes-1, 1.5, 2, and 2.5 on the basis of body weight.…”
Section: Introductionmentioning
confidence: 99%
“…All these modifications suited for easier placement by users who are inexperienced with laryngeal masks. 17 This study shows that mechanical ventilation is satisfactorily maintained with the LMA Supreme™ throughout the surgery.…”
Section: Discussionmentioning
confidence: 60%
“…It has a rigid double-lumen tube with a double aperture design that facilitates the introduction of an orogastric tube to aspirate gastric contents, and it can be inserted without use of introducer devices. In clinical trials, the SLMA has been found comparable with the reusable LMA ProSeal TM (PLMA), 1 with the SLMA having a higher rate of success on first attempt insertion. [2][3][4][5] Previous studies have shown the use of the PLMA and SLMA in non-paralyzed patients, 6 in anesthetized patients with a simulated difficult airway, 7 and in patients undergoing laparoscopy procedures.…”
Section: Résumémentioning