2003
DOI: 10.1016/s0165-9936(03)00301-7
|View full text |Cite
|
Sign up to set email alerts
|

Liquid-sample introduction in plasma spectrometry

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
53
1
1

Year Published

2005
2005
2015
2015

Publication Types

Select...
5
3

Relationship

2
6

Authors

Journals

citations
Cited by 75 publications
(59 citation statements)
references
References 22 publications
(21 reference statements)
0
53
1
1
Order By: Relevance
“…In order to explain this behaviour, it must be considered that the ICP emission signal depends mainly on the characteristics of the aerosols generated by the nebulizer and on the spray chamber design. 1,26 Finer aerosols improve solution (i.e. analyte and solvent) transport to the plasma, and hence the emission signal.…”
Section: Resultsmentioning
confidence: 99%
“…In order to explain this behaviour, it must be considered that the ICP emission signal depends mainly on the characteristics of the aerosols generated by the nebulizer and on the spray chamber design. 1,26 Finer aerosols improve solution (i.e. analyte and solvent) transport to the plasma, and hence the emission signal.…”
Section: Resultsmentioning
confidence: 99%
“…1 Charge transfer reactions from C + species to analyte atoms. 2 Improvement of aerosol transport efficiency. 3 Not discussed.…”
Section: Non-spectroscopic Interferences or Matrix Effects Spectroscomentioning
confidence: 99%
“…Inductively coupled plasma (ICP) techniques, particularly renowned for their sensitivity and selectivity [1], are the most widely used for trace-element measurements [2]. Three ICP technologies have been developed: optical emission spectrometry (ICP-OES, also called atomic emission spectroscopy, ICP-AES), mass spectrometry (ICP-MS) [1,3] and to a lesser extent atomic fluorescence spectrometry (ICP-AFS).ICP sources allow the atomization and the almost complete ionization of all elements in a wide range of samples [4].…”
Section: Introductionmentioning
confidence: 99%
“…Wash out time can be defined as the time required for reaching 1% of the stable signal after blank introduction, 1 and it depends on the experimental conditions and the system design (i.e. inner volume) used.…”
Section: 31mentioning
confidence: 99%
“…1 Liquid samples are usually introduced in ICP-based techniques by means of a pneumatic concentric nebulizer attached to a spray chamber.…”
Section: Introductionmentioning
confidence: 99%