2011
DOI: 10.1097/scs.0b013e318232a734
|View full text |Cite
|
Sign up to set email alerts
|

Life-Threatening Complication of Recurrent Laryngeal Nerve Monitoring With EMG Reinforced Silicone ETT

Abstract: We present 2 cases with complication of herniation of a specially designed endotracheal tube (ETT) cuff used for laryngeal nerve monitoring during thyroid surgery. Abrupt and total blockage of ventilation has occurred at 35th and 40th minutes of general anesthesia. In an animal trachea, we showed that overinflated cuff may herniate and block ETT tip easily. A 25-mm upward movement of the ETT is found as a contributing factor to cuff herniation.In patients intubated with electromyographic reinforced ETT, any ev… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
7
0

Year Published

2013
2013
2021
2021

Publication Types

Select...
8
1

Relationship

0
9

Authors

Journals

citations
Cited by 11 publications
(7 citation statements)
references
References 1 publication
0
7
0
Order By: Relevance
“…In an animal trachea, the same authors showed that overinflated cuff may herniate and block ETT tip easily (12). A 25-mm upward movement of the ETT was found as a contributing factor to cuff herniation.…”
Section: Ett Complicationsmentioning
confidence: 86%
“…In an animal trachea, the same authors showed that overinflated cuff may herniate and block ETT tip easily (12). A 25-mm upward movement of the ETT was found as a contributing factor to cuff herniation.…”
Section: Ett Complicationsmentioning
confidence: 86%
“…A learning curve has been described for their accurate placement, where the EMG endotracheal tube required readjustment for successful recurrent larygneal nerve monitoring after intubation in 10% of the patients (Dionigi et al, 2008). EMG endotracheal tubes may be more prone than other endotracheal tubes to cuff herniation after overfilling and tube traction, resulting in the blockage of ventilation (Oysu and Demir, 2011).…”
Section: Discussionmentioning
confidence: 99%
“…Unlike during thyroid surgery, the cuff electrode cannot be relocated to the true vocal fold during 1-lung ventilation, as it would jeopardize adequate ventilation and interfere with the operative field. Some cases in which the E-tube balloon completely blocked the airway during operation have been reported [ 6 ].…”
Section: Discussionmentioning