2015
DOI: 10.1016/j.radmeas.2014.11.003
|View full text |Cite
|
Sign up to set email alerts
|

LiBe-14: A novel microdosimeter using LiF and BeO thermoluminescence dosimeter pairs for clinical and aerospace applications

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
12
0

Year Published

2016
2016
2021
2021

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 6 publications
(12 citation statements)
references
References 23 publications
0
12
0
Order By: Relevance
“…The commercially available TEPC (Model REM 500B; Supplier Health Physics Instruments, Goleta, CA 93177, U.S.A.) based on a 597 mm diameter spherical Rossi Counter (RC) was used in this work (figure 2). The counter was filled with tissue-equivalent Propane gas at 1.77kPa pressure emulating a 2 µm diameter biological cell volume [8]. Pairs of TLD-700 and BeO were attached to the TEPC and exposed to secondary (lateral) radiation field produced by bombarding a polystyrene plate phantom with protons of various energies (figure 2).…”
Section: Principle Of Tepc and Tld-microdosimetermentioning
confidence: 99%
See 3 more Smart Citations
“…The commercially available TEPC (Model REM 500B; Supplier Health Physics Instruments, Goleta, CA 93177, U.S.A.) based on a 597 mm diameter spherical Rossi Counter (RC) was used in this work (figure 2). The counter was filled with tissue-equivalent Propane gas at 1.77kPa pressure emulating a 2 µm diameter biological cell volume [8]. Pairs of TLD-700 and BeO were attached to the TEPC and exposed to secondary (lateral) radiation field produced by bombarding a polystyrene plate phantom with protons of various energies (figure 2).…”
Section: Principle Of Tepc and Tld-microdosimetermentioning
confidence: 99%
“…Hence, BeO and 7 LiF: Mg,Ti, (TLD-700) pairs cross-calibrated using a TEPC could be used as a passive microdosimeter, i.e. surrogate of a much larger gas-filled TEPC [8,13].…”
Section: Principle Of Tepc and Tld-microdosimetermentioning
confidence: 99%
See 2 more Smart Citations
“…(3) Since then, there have been many reports on the properties and applications of thermoluminescent phosphors based on BeO. (4)(5)(6)(7)(8) However, there have been only a few reports on TL glow component analysis, despite this being necessary to understand the physical characteristics of TL based on BeO. Nieto et al reported studies on the kinetic parameters of undoped BeO using an isothermal decay method, for which the calculated activation energy (E) was in agreement with that obtained using the TL glow curve shape with Chen's modified method; however, the frequency factor (s) differed considerably due to the inclusion of an experimental error.…”
Section: Introductionmentioning
confidence: 99%