Left ventricular sensing in cardiac resynchronization devices—opportunities and pitfalls for device programming

Abstract: Introduction: Some cardiac resynchronization therapy (CRT) device manufacturers (Biotronik, Germany; Boston Scientific, United States) have implemented left ventricular (LV) sensing functionality to prevent pacing into the vulnerable phase.Physicians are only partially aware of programming pitfalls related to LV sensing and general programming advice is lacking. Methods and Results:We provide an illustrative case-series-based review of the variety of potential problems with LV sensing. LV sensing may inappropr… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
16
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 7 publications
(16 citation statements)
references
References 18 publications
0
16
0
Order By: Relevance
“…The programmed duration of the LVURI or LVPP plays a crucial role in this situation and has a deleterious effect on biventricular pacing (Figure 4). The appearance of lock‐in in devices from Boston Scientific seems unlikely, as the sensing in these devices is coupled with a nonprogrammable LV refractory period of at least 250 ms 12 . This refractory period would cover even long interventricular delays (RVp‐to‐LVs), preventing sustained lock‐in.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…The programmed duration of the LVURI or LVPP plays a crucial role in this situation and has a deleterious effect on biventricular pacing (Figure 4). The appearance of lock‐in in devices from Boston Scientific seems unlikely, as the sensing in these devices is coupled with a nonprogrammable LV refractory period of at least 250 ms 12 . This refractory period would cover even long interventricular delays (RVp‐to‐LVs), preventing sustained lock‐in.…”
Section: Discussionmentioning
confidence: 99%
“…However, LV sensing is not only used to monitor the LV lead but is also associated with LV pacing inhibition (ie, to prevent pacing during the vulnerable phase of the cardiac cycle). This can reduce the amount of biventricular pacing and even trigger arrhythmias 10‐12 …”
Section: Introductionmentioning
confidence: 99%
“…BIOTRONIK and Boston Scientific devices used for CRT can sense from the LV lead . LV sensing was designed primarily to prevent competitive pacing outside the LV myocardial absolute refractory period and it works by inhibiting the release of an LV stimulus in the vulnerable period of the LV myocardium.…”
Section: Discussionmentioning
confidence: 99%
“…LV sensing in Boston Scientific and BIOTRONIK CRT devices do not influence traditional DDD RV timing cycles. The special LV sensing function is based on limitation of the LV pacing rate by means of the LVURI . The LV Protection Period of Boston Scientific devices (LVPP) is equivalent to the LVURI .…”
Section: Discussionmentioning
confidence: 99%