Late failure of a Nellix endoprosthesis treated with the t-Branch off-the-shelf multibranched stent graft

Abstract: A 66-year-old man arrived at our emergency department 3 years after an endovascular aneurysm sealing procedure performed at another center. Computed tomography angiography showed distal migration of a Nellix endoprosthesis (Endologix, Irvine, Calif) and a posterior contained rupture. The left kidney was excluded by an occluded iliac-left renal bypass, which was performed at the time of the Nellix implantation because of unintended coverage. A t-Branch (Cook Medical, Bloomington, Ind) was implanted in an emerge… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(1 citation statement)
references
References 14 publications
(15 reference statements)
0
1
0
Order By: Relevance
“…Therefore, in each case, we performed a predilatation of the ipsilateral Nellix with a 12 × 60mm balloon. Lorido et al described the use of the off-the-shelf t-Branch stent-graft in a patient with an occluded iliacrenal bypass [21]. Although the stent-graft is delivered with the 22-F introduction system, the authors describe the damage of the endoskeleton of the Nellix device despite its predilatation.…”
Section: Discussionmentioning
confidence: 99%
“…Therefore, in each case, we performed a predilatation of the ipsilateral Nellix with a 12 × 60mm balloon. Lorido et al described the use of the off-the-shelf t-Branch stent-graft in a patient with an occluded iliacrenal bypass [21]. Although the stent-graft is delivered with the 22-F introduction system, the authors describe the damage of the endoskeleton of the Nellix device despite its predilatation.…”
Section: Discussionmentioning