Laser etching causing fatigue fracture at the neck–shoulder junction of an uncemented femoral stem: A case report

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

0
9
0

Year Published

2015
2015
2023
2023

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 15 publications
(10 citation statements)
references
References 9 publications
(3 reference statements)
0
9
0
Order By: Relevance
“…Extractions of such broken hip stems during revision surgery are extensive and may require Cases of implant fatigue due to laser etching were previously reported where Woolson et al [20] showed fractures of cemented cobalt chromium stem components. Furthermore, fractures in cobalt-chromium and titanium alloyed hip implants at the laser etched sites of femoral necks and stems were published [8,18,21]. To our knowledge, this is the first study of femoral stem fracture of TMZF alloy hip implants due to laser etching.…”
Section: Discussionmentioning
confidence: 92%
“…Extractions of such broken hip stems during revision surgery are extensive and may require Cases of implant fatigue due to laser etching were previously reported where Woolson et al [20] showed fractures of cemented cobalt chromium stem components. Furthermore, fractures in cobalt-chromium and titanium alloyed hip implants at the laser etched sites of femoral necks and stems were published [8,18,21]. To our knowledge, this is the first study of femoral stem fracture of TMZF alloy hip implants due to laser etching.…”
Section: Discussionmentioning
confidence: 92%
“…Therefore, the radius of the notch is, as a geometrical function, important [42]. Jang et al demonstrated that thermal damage of laser etching on hip implants led to their early fracture [43]. Huber et al [25] and Konrads et al [26] reported a decrease in the fatigue life of Ti hip implants due to the thermal damage of electrocautery.…”
Section: Discussionmentioning
confidence: 99%
“…Several other risk factors for stem breakage have previously been identified including laser etching of the stem and other potential mechanical stress risers. 14,[42][43][44][45] A further commonly cited risk factor is component varus positioning. 15,39,46 While the majority of cemented stem breakages reported in the literature are in the mid-third region, there are some reports of cemented stems breaking through the neck.…”
Section: Discussionmentioning