Abstract:The 500 microF, long duration waveform reduced peak voltage 41% (374 +/- 18 V [STD] vs 219 +/- 14 V [LD], P < 0.001) and reduced peak current 38% (11.0 +/- 1.1 A [STD] vs 6.8 +/- 0.6 A [LD], P < 0.001) but did not significantly change the delivered energy (12.4 +/- 1.3 J [STD] vs 13.4 +/- 1.0 J [LD]). Durations increased from 10.0 +/- 0.2 to 17.6 +/- 0.5 msec (P < 0.001).(ABSTRACT TRUNCATED AT 250 WORDS)
“…We chose the 300-F capacitor because it has been shown that large-capacitor waveforms reduce DFT peak voltage without increasing DFT energy. 29,30 Reducing peak voltage may decrease detrimental shock effects and may allow lower voltage ratings of the electronic components in the defibrillator, which would allow the size of the components to be reduced.…”
Background
—It has been reported that triphasic defibrillation waveforms cause less myocardial injury than biphasic waveforms. This study compared the defibrillation thresholds (DFTs) of triphasic and biphasic waveforms.
Methods and Results
—DFTs were determined for a transvenous lead system and a 300-μF-capacitor defibrillator. In 8 pigs (group 1), DFTs were determined for 5 triphasic waveforms with tilts of 80%, 83%, and 86% and for 1 biphasic waveform. DFTs were determined in another 8 pigs (group 2) for 2 triphasic and 4 biphasic waveforms with tilts of 43%, 49%, and 56%. In both groups, a biphasic waveform from a 140-μF-capacitor defibrillator was also evaluated, and both shock polarities were tested for each waveform. In group 1, with the 300-μF-capacitor defibrillator, the leading-edge voltage and energy stored at DFT were significantly lower for triphasic waveforms with phase-duration ratios of 50/33/17 and an anode at the right ventricular electrode for phase 1 than for biphasic waveforms (
P
<0.001). In group 2, the stored energy of triphasic waveforms with 56% and 49% tilt was significantly lower than that of biphasic waveforms with the same tilts for anodal but not cathodal phase 1 at the right ventricular electrode. Electrode polarity significantly affected the DFT of triphasic waveforms for both studies.
Conclusions
—Some 80% tilt triphasic waveforms defibrillate more efficiently than biphasic waveforms with a 300-μF-capacitor defibrillator. The triphasic waveforms for both groups were not superior to 140-μF-capacitor biphasic waveforms. The efficacy of triphasic waveforms depends on phase durations and electrode polarity.
“…We chose the 300-F capacitor because it has been shown that large-capacitor waveforms reduce DFT peak voltage without increasing DFT energy. 29,30 Reducing peak voltage may decrease detrimental shock effects and may allow lower voltage ratings of the electronic components in the defibrillator, which would allow the size of the components to be reduced.…”
Background
—It has been reported that triphasic defibrillation waveforms cause less myocardial injury than biphasic waveforms. This study compared the defibrillation thresholds (DFTs) of triphasic and biphasic waveforms.
Methods and Results
—DFTs were determined for a transvenous lead system and a 300-μF-capacitor defibrillator. In 8 pigs (group 1), DFTs were determined for 5 triphasic waveforms with tilts of 80%, 83%, and 86% and for 1 biphasic waveform. DFTs were determined in another 8 pigs (group 2) for 2 triphasic and 4 biphasic waveforms with tilts of 43%, 49%, and 56%. In both groups, a biphasic waveform from a 140-μF-capacitor defibrillator was also evaluated, and both shock polarities were tested for each waveform. In group 1, with the 300-μF-capacitor defibrillator, the leading-edge voltage and energy stored at DFT were significantly lower for triphasic waveforms with phase-duration ratios of 50/33/17 and an anode at the right ventricular electrode for phase 1 than for biphasic waveforms (
P
<0.001). In group 2, the stored energy of triphasic waveforms with 56% and 49% tilt was significantly lower than that of biphasic waveforms with the same tilts for anodal but not cathodal phase 1 at the right ventricular electrode. Electrode polarity significantly affected the DFT of triphasic waveforms for both studies.
Conclusions
—Some 80% tilt triphasic waveforms defibrillate more efficiently than biphasic waveforms with a 300-μF-capacitor defibrillator. The triphasic waveforms for both groups were not superior to 140-μF-capacitor biphasic waveforms. The efficacy of triphasic waveforms depends on phase durations and electrode polarity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.