2010
DOI: 10.1109/tasc.2010.2040374
|View full text |Cite
|
Sign up to set email alerts
|

Iseult/INUMAC Whole Body 11.7 T MRI Magnet Status

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
21
0
1

Year Published

2010
2010
2024
2024

Publication Types

Select...
5
3

Relationship

1
7

Authors

Journals

citations
Cited by 43 publications
(22 citation statements)
references
References 4 publications
0
21
0
1
Order By: Relevance
“…On the basis of that study we finally decided to implement 6 quench heaters regularly distributed on the external coil envelop. The complete coil quench is achieved within a delay of 5.8 s leading to a final average shielding coils temperature of 70 K and, in case of accidental quench, to safe hot spot temperatures in the main coil ( 120 K) and in the shielding coils ( 150 K) [1].…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…On the basis of that study we finally decided to implement 6 quench heaters regularly distributed on the external coil envelop. The complete coil quench is achieved within a delay of 5.8 s leading to a final average shielding coils temperature of 70 K and, in case of accidental quench, to safe hot spot temperatures in the main coil ( 120 K) and in the shielding coils ( 150 K) [1].…”
Section: Discussionmentioning
confidence: 99%
“…1): one wet and cryostable main coil (composed itself of 3 main coils) and two fully impregnated shielding coils [1]. A classical way to protect a magnet, especially fully impregnated magnets, is to use the enthalpy tank available in the magnet itself in addition to an external dump resistor to dissipate the stored magnetic energy.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…Low field superconducting MRI magnets were extensively investigated [4]- [7] in the past, uniquely using Nb-Ti superconducting material and operated in 4.2 K liquid helium temperature, with the exception of a 11.7 T Iseult-Inumac whole body MRI magnet [8], which operated in superfluid helium temperature to utilize enhanced upper critical field of niobium titanium. The 11.7 T Iseult-Inumac MRI magnet is designed by stacking Nb-Ti/Cu double pancakes which is wound with Rutherford flat cables.…”
Section: Introductionmentioning
confidence: 99%
“…I N ORDER TO build Iseult, the first full body 500 MHz MRI system [1]- [3], a dedicated conductor R&D program has been made with Alstom (MSA) to validate the best conductor solution [4]. Iseult is a shielded MRI magnet wound with two NbTi types of conductor: -The main conductor for the pancake solenoid creating the 11.75 T magnetic field.…”
Section: Introductionmentioning
confidence: 99%