Investigation of Scattered Radiation Reduction Effect by Use of Collimator Cover Covered with Leaded Sheet in Angiography Equipment

Abstract: The purpose of this study was to investigate the effect of scattered radiation reduction to medical staff by attaching the leaded sheet on the collimator cover of the angiography equipment. Ambient dose equivalent was measured to compare the rate of scattered radiation reduction between with and without the leaded sheet. Shielding effect was confirmed for scattered radiation in all directions, especially 27% of shielding ratio in the head and neck area when angiography equipment installed with small detector, … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2

Citation Types

0
4
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(4 citation statements)
references
References 8 publications
0
4
0
Order By: Relevance
“…Similar to the work of Miyagawa et al, this work investigated the efficacy of a lead sheet secured to the collimator assembly cover . By comparing this work, Miyagawa et al used a block PMMA scatter phantom and measured total scatter to an ionization chamber positioned to mimic the location of a performing physician.…”
Section: Discussionmentioning
confidence: 99%
See 3 more Smart Citations
“…Similar to the work of Miyagawa et al, this work investigated the efficacy of a lead sheet secured to the collimator assembly cover . By comparing this work, Miyagawa et al used a block PMMA scatter phantom and measured total scatter to an ionization chamber positioned to mimic the location of a performing physician.…”
Section: Discussionmentioning
confidence: 99%
“…Similar to the work of Miyagawa et al, this work investigated the efficacy of a lead sheet secured to the collimator assembly cover. 13 By comparing this work, Miyagawa et al used a block PMMA scatter phantom and measured total scatter to an ionization chamber positioned to mimic the location of a performing physician. They measured scatter at three elevations (40, 80, and 130 cm from the floor), at four positions around the patient table, and using five different projection angles, but did not specifically isolate scatter originating from within the phantom vs the collimator assembly.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations