Inter-laboratory comparison of three earplug fit-test systems

Abstract: The National Institute for Occupational Safety and Health (NIOSH) sponsored tests of three earplug fit-test systems (NIOSH HPD Well-Fit™, Michael & Associates FitCheck, and Honeywell Safety Products VeriPRO®). Each system was compared to laboratory-based real-ear attenuation at threshold (REAT) measurements in a sound field according to ANSI/ASA S12.6-2008 at the NIOSH, Honeywell Safety Products, and Michael & Associates testing laboratories. An identical study was conducted independently at the U.S. Army Aero… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
11
0

Year Published

2019
2019
2023
2023

Publication Types

Select...
7

Relationship

2
5

Authors

Journals

citations
Cited by 13 publications
(11 citation statements)
references
References 18 publications
(20 reference statements)
0
11
0
Order By: Relevance
“…Poorly-fit earplugs can enhance the stimulus signal and result in negative attenuations as well. From an inter-laboratory study that NIOSH conducted assessing the performance of HPD Well-Fit, a range of ±2 dB can be expected for the repeatability of a threshold assessment (Byrne et al 2016). Regardless of the cause, workers with obviously low attenuations were reinstructed and retested.…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…Poorly-fit earplugs can enhance the stimulus signal and result in negative attenuations as well. From an inter-laboratory study that NIOSH conducted assessing the performance of HPD Well-Fit, a range of ±2 dB can be expected for the repeatability of a threshold assessment (Byrne et al 2016). Regardless of the cause, workers with obviously low attenuations were reinstructed and retested.…”
Section: Resultsmentioning
confidence: 99%
“…Over time, other fit-testing systems have been developed and studied for their performance both in laboratory and field environments (Murphy, 2013; Berger et al, 2011; Schulz, 2011; Byrne et al 2016). Microphone in real-ear (MIRE) measurement systems such as Phonak’s SafetyMeter and 3M’s EARfit Validation System require use two microphones measure noise reduction of a protector.…”
Section: Introductionmentioning
confidence: 99%
“…In addition, the 3-freq test protocol has been shown to be in agreement with the gold standard REAT method. [18] On the basis of the current results, there is little rationale to support additional test frequencies beyond the three including in the 3-freq protocol used for this study.…”
Section: Discussionmentioning
confidence: 99%
“…Output from this system type, however, has not been as closely correlated with the other two system types and/or the gold standard method. [18] There are pros and cons to each system and system type. Regardless, the intent of all of these systems is to provide an individual assessment of the performance of an HPD on a given user in the field.…”
Section: Introductionmentioning
confidence: 99%
“…Participants were tested using the Michael and Associates' FitCheck Solo (State College, Pennsylvania) REAT fit-testing system. This software program has been verified against ANSI S12.6-2008 (ANSI 2008) standards (Byrne et al 2017). The program was installed on a laptop and used a mousebased patient switch to measure participants' responses, and a sound-attenuating FitCheck earcup (TDH-49P) connected to the laptop with a 3.5mm electrical connector to deliver audible test stimuli.…”
Section: Fit-testingmentioning
confidence: 99%