Integrated Flexible Electronic Devices Based on Passive Alignment for Physiological Measurement

Abstract: This study proposes a simple method of fabricating flexible electronic devices using a metal template for passive alignment between chip components and an interconnect layer, which enabled efficient alignment with high accuracy. An electrocardiogram (ECG) sensor was fabricated using 20 µm thick polyimide (PI) film as a flexible substrate to demonstrate the feasibility of the proposed method. The interconnect layer was fabricated by a two-step photolithography process and evaporation. After applying solder past… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2017
2017
2017
2017

Publication Types

Select...
1

Relationship

1
0

Authors

Journals

citations
Cited by 1 publication
(1 citation statement)
references
References 20 publications
0
1
0
Order By: Relevance
“…Fabrication of ECG Sensor : A flexible ECG sensor was fabricated using the proposed method . Briefly, a sensor device was fabricated on the 20 µm thick polyimide film using two‐step photolithography and passive alignment.…”
Section: Methodsmentioning
confidence: 99%
“…Fabrication of ECG Sensor : A flexible ECG sensor was fabricated using the proposed method . Briefly, a sensor device was fabricated on the 20 µm thick polyimide film using two‐step photolithography and passive alignment.…”
Section: Methodsmentioning