Insensible Water Loss During Extracorporeal Membrane Oxygenation: An In Vitro Study

Abstract: To measure insensible fluid loss from silicone membrane oxygenators during extracorporeal membrane oxygenation (ECMO), an in vitro system was used. A standard neonatal ECMO circuit (Avecor) was connected to a noncompliant reservoir, which was then primed with normal saline. The experiment was conducted by using two silicone oxygenators (Avecor 0.4 and 0.8 m2), three gas flow rates (0.5, 1.0, and 2.0 L/min) (sweep), and two fluid flow rates (200 and 400 ml/min). Two methods were used to measure the water loss. … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
18
0

Year Published

2006
2006
2020
2020

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 9 publications
(20 citation statements)
references
References 4 publications
2
18
0
Order By: Relevance
“…Camacho, et al showed that silicone membranes, historically used for ECMO, had evaporative water loss that correlated with gas flow rate across the membrane surface. 12 Membrane size, transmembrane pressure gradient, and pump flow rate were shown not to be factors of water loss in this study. The silicone membrane is the only oxygenation device currently approved by the FDA for long-term use.…”
Section: Discussionmentioning
confidence: 45%
See 1 more Smart Citation
“…Camacho, et al showed that silicone membranes, historically used for ECMO, had evaporative water loss that correlated with gas flow rate across the membrane surface. 12 Membrane size, transmembrane pressure gradient, and pump flow rate were shown not to be factors of water loss in this study. The silicone membrane is the only oxygenation device currently approved by the FDA for long-term use.…”
Section: Discussionmentioning
confidence: 45%
“…Our data are similar to data from the Camacho, et al study, suggesting that membrane types without micro-porous surfaces yield a somewhat lower insensible water loss value than that of micro-porous membrane oxygenators ( Figure 3). 6,12,13 In conclusion, the Jostra Quadrox D oxygenator has lower insensible water loss per liter of sweep gas than other types of oxygenators described in the recent literature. This study provides useful data for management of fluid balance in patients of all sizes who need ECMO support in the critical care setting.…”
Section: Discussionmentioning
confidence: 70%
“…4 We were unable to explain why such a difference between the two PMP membrane devices existed. Eventually, we ran a pilot study, comparing the Quadrox D and the Hilite ® 2400LT.…”
Section: Discussionmentioning
confidence: 90%
“…The volume of actual water loss has been shown to vary between individual devices and the particular membranes used within the device. Camacho et al, 4 using a silicone membrane oxygenator, reported a 72-ml water loss over a 24-hour period, per litre/minute (L/min) of gas flow. Alexander et al 5 studied a microporous, hollow-fibre, membrane oxygenator and reported an 83-ml water loss over a 24 hour period, per L/min gas flow.…”
Section: Introductionmentioning
confidence: 99%
“…We also chose our methods on the basis of the previous reports of crystalloid to assess in vitro membrane performance. 9,10 It should be noted that the fluid loss across a membrane may vary over time. The initial 72-hour run suggests that the evaporative fluid loss remains constant over an extended period.…”
Section: Study Limitationsmentioning
confidence: 99%