2001
DOI: 10.1159/000052477
|View full text |Cite
|
Sign up to set email alerts
|

Initial Experience with a New Transurethral Microwave Thermotherapy Treatment Protocol ‘30–Minute TUMT’

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
5
0

Year Published

2003
2003
2011
2011

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(6 citation statements)
references
References 16 publications
(17 reference statements)
1
5
0
Order By: Relevance
“…After 3 months of follow-up, a mean Q max value of 17.2 ml/s was reported, compared to a mean 10.4 ml/s before treatment. The Q max increase after '30-min TUMT', compared to older Prostasoft 2.5 software, is similar to that obtained after TURP mostly after 6-12 months, as preliminary results in the literature seem to confirm [27]. Ninety-five percent of treated subjects demonstrated cavitation inside the prostatic adenoma.…”
Section: Discussionsupporting
confidence: 77%
“…After 3 months of follow-up, a mean Q max value of 17.2 ml/s was reported, compared to a mean 10.4 ml/s before treatment. The Q max increase after '30-min TUMT', compared to older Prostasoft 2.5 software, is similar to that obtained after TURP mostly after 6-12 months, as preliminary results in the literature seem to confirm [27]. Ninety-five percent of treated subjects demonstrated cavitation inside the prostatic adenoma.…”
Section: Discussionsupporting
confidence: 77%
“…Tolerability of this new high-energy protocol is high, in our opinion, and it is better than the old protocol, 2.5, because of the shorter treatment time [8,9].…”
Section: Discussionmentioning
confidence: 77%
“…The system for TUMT consists of a microwave generator, a urethral treatment catheter with a microwave antenna and a fiberoptic temperature monitoring system (accuracy B 0.1°C) inside the catheter and in a rectal probe [8,9]. Computer software with 30 min duration and a maximum microwave energy of 80 W controls the microwave generator and the cooling system in response to preset temperatures of the rectal and urethral thermosensors.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations