2018
DOI: 10.1161/circinterventions.117.005923
|View full text |Cite
|
Sign up to set email alerts
|

Infolding of the Venus P-Valve After Transcatheter Pulmonary Valve Implantation

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

1
5
0

Year Published

2018
2018
2024
2024

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 10 publications
(7 citation statements)
references
References 4 publications
1
5
0
Order By: Relevance
“…Continued surveillance is warranted to look for late mechanical disturbances in the valve frame. [ 32 ]…”
Section: Discussionmentioning
confidence: 99%
“…Continued surveillance is warranted to look for late mechanical disturbances in the valve frame. [ 32 ]…”
Section: Discussionmentioning
confidence: 99%
“…More than 100 Venus P valves have been implanted worldwide. 7 The initial experience with this valve is promising with excellent patient outcomes. 8 The reported complications after Venus P valve placement were mainly stent migration and stent fractures.…”
Section: Discussionmentioning
confidence: 99%
“…Patients at risk for infolding have marked eccentricity or eccentric calcification of the annulus, especially with valve oversizing, mostly in large valves (29 mm or 34 mm), and resheathing has been attempted. Infolding has been reported with other self‐expanding nitinol valves, such as the Venus P‐Valve for transcatheter pulmonary valve implantation and with self‐expanding nitinol endovascular stent grafts for abdominal aorta and thoracic aorta aneurysms.…”
Section: Discussionmentioning
confidence: 99%