2005
DOI: 10.1016/j.ijms.2005.09.004
|View full text |Cite
|
Sign up to set email alerts
|

Impulsional mode operation for a Paul ion trap

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
7
0

Year Published

2006
2006
2015
2015

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 11 publications
(7 citation statements)
references
References 8 publications
0
7
0
Order By: Relevance
“…Hence, it is important to know that β z point are the equivalent points; two operating points located in their corresponding stability diagram have the same β z . 31 For the same 0< β z <1 we have, 0<…”
Section: Phase Space Ion Trajectorymentioning
confidence: 93%
See 1 more Smart Citation
“…Hence, it is important to know that β z point are the equivalent points; two operating points located in their corresponding stability diagram have the same β z . 31 For the same 0< β z <1 we have, 0<…”
Section: Phase Space Ion Trajectorymentioning
confidence: 93%
“…26,30 To compute the accurate elements of the motion ----------------- ------------------- ------------------sh c 2 λ i ( ) -------------------⋅ ------------------ch c 2 λ i ( ) ------------------sh c 2 λ i ( ) = equations for the stability diagrams, we have used the fifth order Runge-Kutta numerical method with a 0.001 steps increment for Matlab software and scanning method. Figure 1 (a) and (b) shows the calculated first and second stability regions for the quadrupole ion trap and cylindrical ion trap, 31 black line (solid line): QIT and blue line (dash line): CIT with optimum radius size z 1 = 1.04978z 0 , (a): first stability region and (b): second stability region. Figure 1 shows that the apex of the stability parameters a z stayed the same and the apex of the stability parameters q z decrease for CIT to compare with QIT.…”
Section: Stability Regionsmentioning
confidence: 99%
“…The resolution of a quadrupole ion trap 29 and cylindrical ion trap mass spectrometry in general, is a function of the mechanical accuracy of the hyperboloid of the QIT Dr 0 , and the cylindrical of the CIT Dr 1 , and the stability performances of the electronics device such as, variations in voltage amplitude DV, the rf frequency DU 20 and Dk (0 # k < 1), which tell us how accurate the form of the voltage signal is. The factor Dk plays an important role in building the stability diagrams for the purpose of the mass resolution.…”
Section: Fractional Resolutionmentioning
confidence: 99%
“…Quadrupole mass spectrometry has developed through several stages to its current stage of relatively high performance and increasing popularity. 1,2 The quadrupole mass filter principle, which was invented by Paul and Steinwedel 3 has been widely applied to mass spectrometry, [1][2][3][4][5][6][7][8][9][10] ion cooling and spectroscopy, [11][12][13] frequency standards 14,15 and, quantum computing. 16,17 To apply various objectives, various geometries of the mass spectrometer have been suggested.…”
Section: Introductionmentioning
confidence: 99%