2020
DOI: 10.1021/jasms.0c00133
|View full text |Cite
|
Sign up to set email alerts
|

Improving Spatial Resolution of a LTQ Orbitrap MALDI Source

Abstract: Here, we present a simple and cost-effective procedure to improve the spatial resolution of the commercial MALDI source of a LTQ Orbitrap. Based in spatial filtering techniques, we demonstrate that, with minimal modifications of the original setup, the system resolution can be pushed forward to <10 μm. The improved system performance is demonstrated by means of MALDI imaging of human colon biopsies.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
11
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
5

Relationship

4
1

Authors

Journals

citations
Cited by 5 publications
(11 citation statements)
references
References 14 publications
0
11
0
Order By: Relevance
“…Histological sections were prepared and analyzed by MALDI-IMS as described by Garate et al Briefly, DAN was used as a matrix for negative-ion detection and deposited with the aid of our in-house designed sublimator . Sections were scanned in negative-ion mode at 10 μm/pixel for nontumoral tissue and 25 μm/pixel for ccRCC sections, using an orbitrap analyzer (MALDI-LTQ-Orbitrap XL, Thermo Fisher, San Jose, CA), equipped with a modified MALDI source …”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…Histological sections were prepared and analyzed by MALDI-IMS as described by Garate et al Briefly, DAN was used as a matrix for negative-ion detection and deposited with the aid of our in-house designed sublimator . Sections were scanned in negative-ion mode at 10 μm/pixel for nontumoral tissue and 25 μm/pixel for ccRCC sections, using an orbitrap analyzer (MALDI-LTQ-Orbitrap XL, Thermo Fisher, San Jose, CA), equipped with a modified MALDI source …”
Section: Methodsmentioning
confidence: 99%
“…32 Sections were scanned in negative-ion mode at 10 μm/pixel for nontumoral tissue and 25 μm/pixel for ccRCC sections, using an orbitrap analyzer (MALDI-LTQ-Orbitrap XL, Thermo Fisher, San Jose, CA), equipped with a modified MALDI source. 33 Data were acquired with a mass resolution of 60 000 at m/z = 400 and a laser energy of ∼10−20 μJ/pulse, depending on the spatial resolution. MALDI-IMS experiments were completed and the matrix was removed.…”
Section: Materials Andmentioning
confidence: 99%
“… 9 The sections were scanned in negative-ion mode using the orbitrap analyzer of a MALDI-LTQ-Orbitrap XL (Thermo Fisher, San Jose, CA, USA), equipped with a modified MALDI source. 27 …”
Section: Materials and Methodsmentioning
confidence: 99%
“…Briefly, DAN was used as the matrix for negative-ion detection and was deposited with the aid of our in-house-designed sublimator . The sections were scanned in negative-ion mode using the orbitrap analyzer of a MALDI-LTQ-Orbitrap XL (Thermo Fisher, San Jose, CA, USA), equipped with a modified MALDI source …”
Section: Materials and Methodsmentioning
confidence: 99%
See 1 more Smart Citation