1995
DOI: 10.1016/0360-3016(94)00495-7
|View full text |Cite
|
Sign up to set email alerts
|

Implications of tissue heterogeneity for radiosurgery in head and neck tumors

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

5
23
0

Year Published

1999
1999
2021
2021

Publication Types

Select...
6
2

Relationship

1
7

Authors

Journals

citations
Cited by 48 publications
(28 citation statements)
references
References 8 publications
5
23
0
Order By: Relevance
“…The shortcoming of PB algorithms to accurately predict dose in low density media has been described previously 27 , 32 , 45 , 46 , 47 , 48 , 49 . In a recent study of 304 irradiations performed at 221 different institutions, Kry et al (49) observed an overestimate of 4.9% in PB algorithms compared to measurement in the Radiological Physics Center (RPC) thorax phantom used for RTOG credentialing; these results agree favorably with those reported here.…”
Section: Discussionsupporting
confidence: 87%
See 1 more Smart Citation
“…The shortcoming of PB algorithms to accurately predict dose in low density media has been described previously 27 , 32 , 45 , 46 , 47 , 48 , 49 . In a recent study of 304 irradiations performed at 221 different institutions, Kry et al (49) observed an overestimate of 4.9% in PB algorithms compared to measurement in the Radiological Physics Center (RPC) thorax phantom used for RTOG credentialing; these results agree favorably with those reported here.…”
Section: Discussionsupporting
confidence: 87%
“…Acquisition of beam data for SRS and SABR can be particularly challenging, however, due to the small size of the fields employed and the partial volume effects associated with many common dosimeters. Further, small photon beams exhibit a loss of lateral electronic equilibrium on the central axis, producing output factors that falloff rapidly for fields below 10 mm in diameter 26 , 27 , 28 , 29 . Centering of the dosimeter in the beam is also challenging, and improper alignment of the beam and detector can introduce significant errors.…”
Section: Methodsmentioning
confidence: 99%
“…cavities (3)(4)(5)(6)(7) . On the other hand, the under dosing that gives rise to the skin-sparing effect could reduce the dose delivered to the rectal wall during treatment of prostate cancer, as reported by Holmes et al (1) .…”
Section: Introductionmentioning
confidence: 99%
“…More accurate algorithms may prove valuable, especially in reirradiation cases of the head and neck with hypofractionated dosing where accurate prediction of dose distribution to conformal target volumes and organs at risk is critical. Where clinically available, more accurate dose calculation algorithms, such as Monte Carlo-based (MC) method, (4,5,6,7) could prove beneficial in such cases in which there exists a delicate balance between side effects and tumor control.…”
Section: Introductionmentioning
confidence: 99%
“…Nevertheless, the authors concluded that both normal tissue inhomogeneity and surgical air cavities on the target coverage were adequately characterized by the conventional pencil beam algorithm when compared to MC-based calculation. The challenges of air-tissue interfaces were characterized by Solberg and colleagues, (5) who undertook phantom measurements with and without air-gap inserts and evaluated the ability of a MC algorithm to account for dose perturbation using a 10 MV beam. In their smaller field geometry, which used a radiosurgical setup with a 1 cm circular tissue diameter and a 3 mm air cavity at 2.6 cm depth, both film and diode measurements suggested a reduction of central axis dose by up to 21% in the area immediately following the cavity.…”
Section: Introductionmentioning
confidence: 99%