Implementing failure mode and effect analysis to improve the safety of volumetric modulated arc therapy for total body irradiation

Tara Gray,
Alexander Antolak,
Saeed Ahmed
et al.

Abstract: Volumetric-modulated arc therapy for total body irradiation (VMAT-TBI) is a novel radiotherapy technique that has been implemented at our institution. The purpose of this work is to investigate possible failure modes (FMs) in the treatment process and to develop a quality control (QC) program for VMAT-TBI following TG-100 guidelines. Methods: We formed a multidisciplinary team to map out the complete treatment process of VMAT-TBI following the AAPM TG-100 guidelines. This process map gives a visual representat… Show more