Implementation of dosimetry equipment and phantoms at the MedAustron light ion beam therapy facility

Abstract: Purpose: To describe the implementation of dosimetry equipment and phantoms into clinical practice of light ion beam therapy facilities. This work covers not only standard dosimetry equipment such as computerized water scanners, films, 2D-array, thimble, and plane parallel ionization chambers, but also dosimetry equipment specifically devoted to the pencil beam scanning delivery technique such as water columns, scintillating screens or multilayer ionization chambers. Method: Advanced acceptance testing procedu… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

6
58
0

Year Published

2018
2018
2024
2024

Publication Types

Select...
7

Relationship

7
0

Authors

Journals

citations
Cited by 29 publications
(64 citation statements)
references
References 40 publications
6
58
0
Order By: Relevance
“…All measurements were carried out in the horizontal beam line of treatment room 3 (IR3) at the MedAustron synchrotron facility using a remotely controlled water phantom MP3‐PL (PTW, Freiburg, Germany). Further details and descriptions of the equipment and calibration procedures are provided in the supplementary materials and literature . The detectors were positioned at the effective point of measurement zitalicref …”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…All measurements were carried out in the horizontal beam line of treatment room 3 (IR3) at the MedAustron synchrotron facility using a remotely controlled water phantom MP3‐PL (PTW, Freiburg, Germany). Further details and descriptions of the equipment and calibration procedures are provided in the supplementary materials and literature . The detectors were positioned at the effective point of measurement zitalicref …”
Section: Methodsmentioning
confidence: 99%
“…Further details and descriptions of the equipment and calibration procedures are provided in the supplementary materials and literature. [20][21][22] The detectors were positioned at the effective point of measurement z ref . 23,24 Lateral dose (core) profiles were measured using the MicroDiamond (MD) detector due to its small sensitive area minimizing volume averaging effects.…”
Section: Introductionmentioning
confidence: 99%
“…Actually, pre‐clinical pencil beam characterization (acceptance and commissioning) and QA procedures require various relative dose measurements as summarized in a recent review on technologies and procedures of clinical dosimetry for scanned ion beam radiotherapy . Commissioning and QA is not only necessary for the beam parameters but also for the equipment itself and an overview of a comprehensive implementation addressing the specific demands of scanned beams is given by Grevillot et al …”
Section: Detectors For Absorbed Dose In Nonreference Conditionsmentioning
confidence: 99%
“…21) A further complication arises when the response of the LAIC is not uniform over its sensitive area. 24 The relative response as a function of lateral offset can be expressed normalized to the average response over the entire area,…”
Section: B Direct Determination Of Dose-area-product In Water In Amentioning
confidence: 99%