Implementation and commissioning of an integrated micro-CT/RT system with computerized independent jaw collimation

Abstract: A set of computer-controlled motorized jaws for a micro-CT∕RT system were constructed with position reliably better than a tenth of a millimeter. The hardware system is ready for image-guided conformal radiotherapy for small animals with capability of respiratory-gated delivery.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
24
0

Year Published

2014
2014
2019
2019

Publication Types

Select...
7

Relationship

4
3

Authors

Journals

citations
Cited by 26 publications
(24 citation statements)
references
References 26 publications
0
24
0
Order By: Relevance
“…MicroCT imaging on the GE eXplore CT120 scanner is performed with tube voltages between 70 and 120 kVp and radiotherapy with the maximum tube voltage of 120 kVp and tube current of 60 mA, using 1–10 cm diameter circular collimation with ~2 Gy min −1 maximum output. The GE eXplore CT/RT140 microCT imaging is performed in similar fashion to the imager of the GE eXplore CT120 scanner (Jensen et al 2013). Radiotherapy is performed at 140 kVp with a 1.0 mm focal spot with a set of computerized independent jaws and the same SID as the eXplore CT120.…”
Section: Methodsmentioning
confidence: 99%
“…MicroCT imaging on the GE eXplore CT120 scanner is performed with tube voltages between 70 and 120 kVp and radiotherapy with the maximum tube voltage of 120 kVp and tube current of 60 mA, using 1–10 cm diameter circular collimation with ~2 Gy min −1 maximum output. The GE eXplore CT/RT140 microCT imaging is performed in similar fashion to the imager of the GE eXplore CT120 scanner (Jensen et al 2013). Radiotherapy is performed at 140 kVp with a 1.0 mm focal spot with a set of computerized independent jaws and the same SID as the eXplore CT120.…”
Section: Methodsmentioning
confidence: 99%
“…The dose and time frame were chosen on the basis of a previously published study of the effect of brain irradiation on normal brain in an immune‐competent mouse strain . Radiation delivery was performed using an in‐house custom micro‐irradiation system (140 kVp, 50 kW) with on‐board image guidance . Anesthetized mice were placed in the tail‐first prone position and the skull was located using on‐board fluoroscopy.…”
Section: Methodsmentioning
confidence: 99%
“…Computerized collimators were positioned to create a 10 × 14‐mm 2 field that encompassed the brain and shielded the remainder of the mouse head and body. The whole brain was irradiated using two equal‐weighted, parallel‐opposed beams for a cumulative dose of 10 Gy, as described and validated previously . All irradiated mice received their full 10‐Gy dose and no irradiated mice showed any outward sign of adverse effects caused by the irradiation treatment.…”
Section: Methodsmentioning
confidence: 99%
“…A similar MicroCT/RT system was designed by the University of Western Ontario and Robarts Research Institute (London, ON, Canada) [162]. Instead of the variable iris aperture collimator, a computerised asymmetric jaw collimator is installed consisting of two orthogonal jaw pairs.…”
Section: Microct/rt Systemmentioning
confidence: 99%