Abstract. LaBr 3 :Ce,CeBr 3 and GAGG:Ce scintillators were investigated and the determined characteristics were compared with those obtained for the well-known and widely used CsI:Tl and NaI:Tl crystals. All the detectors were of the same size of 10 × 10 × 5 mm 3 . The aim of this test study was to single out scintillation detectors most suitable for -ray spectrometry and -ray emission radial profi le measurements in high-temperature plasma experiments. Decay time, energy resolution, non-proportionality and full energy peak detection effi ciency were measured for -ray energies up to 1770 keV. Due to their good energy resolution, short decay time and high detection effi ciency for MeV gamma rays, LaBr 3 :Ce and CeBr 3 scintillators are proposed as the best candidates for use especially under conditions of high count rates, which are expected in the forthcoming DT experiments.