High-Dose-Rate Brachytherapy for Prostate Cancer: Direct in Vivo Dosimetry Is an Important Factor in Increasing the Effectiveness of Treatment and Reducing the Incidence of Complications

Abstract: The technology of direct dosimetry was examined during a session of automated high-power interstitial gamma therapy for prostate cancer with a source of iridium-192 using a MKD-04 scintillation multi-detector dosimeter developed at the JSC ROSATOM. Miniature detectors of up to 5 pieces allow simultaneous measurements in areas of medical interest - the urethra and the rectum at different distances from the intrastats. The technology of direct dosimetry with rigid fixation of detectors is aimed at increasing qua… Show more