Hemodynamic investigation and in vitro evaluation of a novel mixed‐flow blood pump

Abstract: Introduction Ventricular assist devices (VADs) are considered an effective treatment for patients with advanced heart failure, while complications associated with blood damage remain a burden. Structure design innovation has the potential to reduce hemolysis and improve hemocompatibility. Methods In this research, a novel mixed‐flow blood pump that integrates structural features of the axial and centrifugal VADs was proposed. The pump consists of an inducer, a mixed impeller supported by two ceramic pivot bear… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
4
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(4 citation statements)
references
References 40 publications
0
4
0
Order By: Relevance
“…Reynolds stresses were not included in the scalar shear stress, because associated blood damage models have not reached a consensus 29 …”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…Reynolds stresses were not included in the scalar shear stress, because associated blood damage models have not reached a consensus 29 …”
Section: Methodsmentioning
confidence: 99%
“…Reynolds stresses were not included in the scalar shear stress, because associated blood damage models have not reached a consensus. 29 MIH calculation is of great importance in the estimation of potential blood damage. 30 MIH was calculated using a Euler scalar transport approach using the equation 31 :…”
Section: Blood Damage Estimationmentioning
confidence: 99%
“…For numerical simulations, blood was considered a single-phase flow and an incompressible Newtonian fluid with a density of 1050 kg/m 3 and a viscosity of 0.0035 Pa • s. 11 To analyze the relationship between pressure and flow rate of the blood pump at a steady state, the turbulence model was used the k-epsilon model, the solution method was SIMPLEC, and the wall function was used the standard wall function.…”
Section: Cfd Calculation Model and Parameter Settingmentioning
confidence: 99%
“…However, the effectiveness of this type of device for treating HFpEF has not been extensively studied. Design and analysis studies are needed to evaluate its clinical, biological, and hemodynamic effects and to ensure the reliability of this medical device [ 19 , 20 , 21 , 22 ]. If a centrifugal pump is chosen, it will operate at off-design conditions, far from its nominal point, due to the low flow rate required.…”
Section: Introductionmentioning
confidence: 99%