Space Telescopes and Instrumentation 2022: Ultraviolet to Gamma Ray 2022
DOI: 10.1117/12.2627632
|View full text |Cite
|
Sign up to set email alerts
|

Ground test results of the electromagnetic interference in the x-ray microcalorimeter onboard XRISM

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
2

Relationship

1
1

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 0 publications
0
2
0
Order By: Relevance
“…We found no significant changes in the Mn Kα energy resolution among them. We also performed a long (>8 hr) Fe55 x-ray integration several times by operating the RW1–4 at a fixed rate of 3 kprm and the magnetic torqer, which is a known electromagnetic interference source, 20 at some fixed duty ratio. Comparison with a reference measurement of no operation of these interference sources showed no significant difference in the performance of individual pixels.…”
Section: Resultsmentioning
confidence: 99%
“…We found no significant changes in the Mn Kα energy resolution among them. We also performed a long (>8 hr) Fe55 x-ray integration several times by operating the RW1–4 at a fixed rate of 3 kprm and the magnetic torqer, which is a known electromagnetic interference source, 20 at some fixed duty ratio. Comparison with a reference measurement of no operation of these interference sources showed no significant difference in the performance of individual pixels.…”
Section: Resultsmentioning
confidence: 99%
“…This work used computational resources of the supercomputer Fugaku provided by RIKEN. This paper is a derivation of the SPIE proceeding 33 …”
Section: Acknowledgmentsmentioning
confidence: 99%