2018
DOI: 10.1016/j.numecd.2017.10.023
|View full text |Cite
|
Sign up to set email alerts
|

FreeStyle Libre and Dexcom G4 Platinum sensors: Accuracy comparisons during two weeks of home use and use during experimentally induced glucose excursions

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

4
39
4

Year Published

2018
2018
2022
2022

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 54 publications
(47 citation statements)
references
References 18 publications
4
39
4
Order By: Relevance
“…In contrast, the median ARD was optimal during hyperglycemia. The present study findings are in line with the recently published accuracy of FSL CGMS [14][15][16][17] . The MARD between FSL and Dexcom G4 Platinum (Dexcom Inc., San Diego, CA, USA) CGMS was higher in hypoglycemia than in either euglycemia or hyperglycemia 15,17 .…”
Section: Discussionsupporting
confidence: 93%
See 2 more Smart Citations
“…In contrast, the median ARD was optimal during hyperglycemia. The present study findings are in line with the recently published accuracy of FSL CGMS [14][15][16][17] . The MARD between FSL and Dexcom G4 Platinum (Dexcom Inc., San Diego, CA, USA) CGMS was higher in hypoglycemia than in either euglycemia or hyperglycemia 15,17 .…”
Section: Discussionsupporting
confidence: 93%
“…The present study findings are in line with the recently published accuracy of FSL CGMS [14][15][16][17] . The MARD between FSL and Dexcom G4 Platinum (Dexcom Inc., San Diego, CA, USA) CGMS was higher in hypoglycemia than in either euglycemia or hyperglycemia 15,17 . The MARD of several devices, including the FSL, compared with venous or capillary BG are also higher in hypoglycemia than in euglycemia or hyperglycemia 14,16,17 .…”
Section: Discussionsupporting
confidence: 93%
See 1 more Smart Citation
“…1), which were connected to the leads of a CHI 1440 potentiostat (CH Instruments) to provide 1 Hz measurements of sensor current. The Dexcom G4 is designed for 7 days of use but is often used for up to 14 days (21). Therefore, in vivo experiments were performed 7-13 days after sensor implantation to reflect clinically relevant conditions.…”
Section: Sensorsmentioning
confidence: 99%
“…Direct accuracy comparisons between different CGM systems are uncommon and typically require patients to wear two [9][10][11][12][13] or three [14][15][16] sensors simultaneously and submit to frequent blood sampling. Because most clinical trials evaluating outcomes associated with CGM use blinded CGM or SMBG regimens as the control group, direct comparisons of outcomes associated with different CGM systems are even less common.…”
Section: Introductionmentioning
confidence: 99%