1991
DOI: 10.1259/0007-1285-64-759-228
|View full text |Cite
|
Sign up to set email alerts
|

Fracturing and stretching of angioplasty balloon catheters under bending with tension

Abstract: Eight catheters from five manufacturers have been subjected to tests which combined bending and tension. Three failure mechanisms have been identified. These included separation of the balloon from its proximal attachment to the shaft, fracture of the shaft at its junction with the balloon, and undue ductility of the shaft. It is concluded that good catheter design should ensure that there is no sudden change in stiffness along the length of the catheter, and that the optimum relationship of ductility to stiff… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...

Citation Types

0
0
0

Publication Types

Select...

Relationship

0
0

Authors

Journals

citations
Cited by 0 publications
references
References 10 publications
(11 reference statements)
0
0
0
Order By: Relevance

No citations

Set email alert for when this publication receives citations?