2006
DOI: 10.1097/01.bot.0000199119.38359.96
|View full text |Cite
|
Sign up to set email alerts
|

Fixation of Periprosthetic Femoral Shaft Fractures Associated With Cemented Femoral Stems

Abstract: Locked plating constructs were stiffer than the Ogden construct in axial loading and torsion. Although no differences in loads to failure during torsion were noted, locked plating constructs exhibited catastrophic failure not observed with the Ogden construct.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

1
82
1

Year Published

2008
2008
2022
2022

Publication Types

Select...
7
2

Relationship

0
9

Authors

Journals

citations
Cited by 146 publications
(84 citation statements)
references
References 21 publications
1
82
1
Order By: Relevance
“…This problem is illustrated in figure. 5 . Different fixation methods were attempted and were frequently followed by unsatisfactory results [17] . In many studies, good results were reported with interlocked supracondylar nails if the implant allowed such an option ( fig.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…This problem is illustrated in figure. 5 . Different fixation methods were attempted and were frequently followed by unsatisfactory results [17] . In many studies, good results were reported with interlocked supracondylar nails if the implant allowed such an option ( fig.…”
Section: Discussionmentioning
confidence: 99%
“…Fixation methods include simple screws, cerclage wires, cable systems, ordinary plates alone or combined with strut graft and interlocking nails [10][11][12][13][14][15][16] . Recently developed locked plate systems are popular and very useful for periprosthetic fracture fixation [17,18] . In this report, we present our experience with the treatment of periprosthetic fracture of the femur after hip and knee replacements.…”
Section: Introductionmentioning
confidence: 99%
“…Locking plates demonstrate significantly increased stiffness as a biomechanical construct, however they exhibit catastrophic failure in terms of linear fracture between adjacent screw holes on the proximal cortex in cadaveric models, which is not observed with a cable plate construct [9]. Clinically, no benefit over conventional methods has been shown, with one series of fourteen patients experiencing six failures, with three plate fractures and three instances of fixation pull-out [10].…”
Section: Introductionmentioning
confidence: 99%
“…As a result, placement with abutment of the prosthesis stem leading to altered screw threads with subsequent screw pullout or fatigue failure of the plate have been described [6][7][8]. Moreover, proximal screws may violate the bone-prosthesis interface, lead to formation of cement cracks with subsequent prosthetic loosening, act as stress enhancers with an increasing risk for further fractures, and cerclage fixation has shown high failure rates with minimal rotational stability [9][10][11][12]. For these reasons plates for extracortical fixation using the clamp-on principle have been developed.…”
Section: Introductionmentioning
confidence: 99%