Final report on CCT-K6: Comparison of local realisations of dew-point temperature scales in the range −50 °C to +20 °C

Abstract: A key comparison in dew-point temperature was carried out among the national standards held by NPL (pilot), NMIJ, INTA, VSL, INRIM, MIKES, NIST, NIM, VNIIFTRI-ESB and NMC. A pair of condensation-principle dew-point hygrometers was circulated and used to compare the local realisations of dew point for participant humidity generators in the range −50 °C to +20 °C. The duration of the comparison was prolonged by numerous problems with the hygrometers, requiring some repairs, and several additional check measureme… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
8
0

Year Published

2016
2016
2018
2018

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(8 citation statements)
references
References 0 publications
0
8
0
Order By: Relevance
“…Here, e ( T DP/FP )= e w ( T DP ) for T DP/FP ≥ 0, where e w is the saturated vapor pressure for water, calculated using [12]. Also, e ( T DP/FP )= e i ( T FP ) for T DP/FP < 0, where e i is the saturated vapor pressure for ice, calculated using [5–6].…”
Section: Generatorsmentioning
confidence: 99%
See 2 more Smart Citations
“…Here, e ( T DP/FP )= e w ( T DP ) for T DP/FP ≥ 0, where e w is the saturated vapor pressure for water, calculated using [12]. Also, e ( T DP/FP )= e i ( T FP ) for T DP/FP < 0, where e i is the saturated vapor pressure for ice, calculated using [5–6].…”
Section: Generatorsmentioning
confidence: 99%
“…Because NIST participated in the CCT-K6 multilateral key comparison, some of the results of this bilateral comparison may be linked to the key comparison reference value (KCRV) for T DP/FP [1]. The degree of equivalence between T DP/FP realized by a NMI and the KCRV, D NMI/KCRV , is defined as DNMI/KCRVfalse(TDP/FPfalse)[ΔTDP/FP]NMI[ΔTDP/FP]KCRV.…”
Section: Linkage To the Cct-k6 Kcrvmentioning
confidence: 99%
See 1 more Smart Citation
“…For T DP/FP ≥ −15 °C, the HHG operates as a conventional two-pressure generator, saturating air with water at a temperature T s and pressure P s to produce moist air with a molar fraction x g given by 1) Here, e(T s ) is the water vapor pressure at T s, calculated using [2][3] and f(T s , P s ) is the watervapor enhancement factor, calculated using [4]. The saturator temperature is measured by a standard platinum resistance thermometer (SPRT) immersed in the same temperaturecontrolled bath as the saturator.…”
Section: Generatorsmentioning
confidence: 99%
“…Because NIST participated in the CCT-K6 multilateral key comparison, some of the results of this bilateral comparison may be linked to the key comparison reference value (KCRV) for T DP/FP [1]. The degree of equivalence between T DP/FP realized by a NMI and the KCRV,D NMI/KCRV , is defined as…”
Section: Linkage To the Cct-k6 Kcrvmentioning
confidence: 99%