2016
DOI: 10.5114/jcb.2016.61828
|View full text |Cite
|
Sign up to set email alerts
|

Film based verification of calculation algorithms used for brachytherapy planning-getting ready for upcoming challenges of MBDCA

Abstract: PurposeWell-known defect of TG-43 based algorithms used in brachytherapy is a lack of information about interaction cross-sections, which are determined not only by electron density but also by atomic number. TG-186 recommendations with using of MBDCA (model-based dose calculation algorithm), accurate tissues segmentation, and the structure's elemental composition continue to create difficulties in brachytherapy dosimetry. For the clinical use of new algorithms, it is necessary to introduce reliable and repeat… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
16
0

Year Published

2017
2017
2023
2023

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 7 publications
(16 citation statements)
references
References 31 publications
0
16
0
Order By: Relevance
“…Table 1 highlights the need for developing a single comprehensive template to cover all the required items. Also, several studies argue for the importance of developing a comprehensive template for commissioning 3D-image-based TPSs in HDR GYN BT [4,26,[28][29][30]51,58]. In particular, this paper presents the first comprehensive template for verifying optimization routines and DVH-based parameters in BT TPSs.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Table 1 highlights the need for developing a single comprehensive template to cover all the required items. Also, several studies argue for the importance of developing a comprehensive template for commissioning 3D-image-based TPSs in HDR GYN BT [4,26,[28][29][30]51,58]. In particular, this paper presents the first comprehensive template for verifying optimization routines and DVH-based parameters in BT TPSs.…”
Section: Discussionmentioning
confidence: 99%
“…Inhomogeneity modelling accuracy: Verifies that the TPS models the inhomogeneities in GYN cases and modifies the dose distributions according to the attenuation/scattering properties of these inhomogeneities (especially applicators and shields) [10,19,26,27,30,[53][54][55]. This test is particularly important for the applicators and shields in GYN cases.…”
Section: Model-based Tpss Consistency With the Tg-43 Calculationsmentioning
confidence: 92%
See 1 more Smart Citation
“…Then the calibration curve was prepared and used recalculate the optical density (analog to digital conversion value; ADC) to the doses. 17 …”
Section: Methodsmentioning
confidence: 99%
“…"Static" IMBT approaches include those in which the shield does not move relative to the source or surrounding tissues during BT delivery, such as shielded applicators (eg, shielded cylinders, [3][4][5][6][7][8][9][10][11] shielded tandem and ovoids, [12][13][14][15][16][17][18][19][20] Strut-Adjusted Volume Implants [SAVI], 21 intracavitary mold applicators [ICMA], [22][23][24][25][26] direction-modulated brachytherapy [DirMBT], 21,25,[27][28][29][30][31][32][33][34] and cup or D-shaped applicators [35][36][37][38][39][40][41][42][43][44][45][46][47][48][49]…”
Section: Introductionmentioning
confidence: 99%