2021
DOI: 10.1080/00015385.2021.1939510
|View full text |Cite
|
Sign up to set email alerts
|

FFR pressure wire comparative study: piezoresistive versus optical sensor

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(6 citation statements)
references
References 10 publications
0
3
0
Order By: Relevance
“…Previous studies of the relationship between IMR and MVO have shown that the IMR is higher in patients with STEMI and MVO ( 14 16 ). However, IMR is typically measured using a thermodilution wire with a sensor near its tip, which has less torquability in comparison with workhorse wires, leading to technical difficulty and increased cost ( 17 ). Furthermore, IMR requires pharmacological induction of hyperemia, which may be uncomfortable for patients and has a risk of morbidity from arrhythmia.…”
Section: Discussionmentioning
confidence: 99%
“…Previous studies of the relationship between IMR and MVO have shown that the IMR is higher in patients with STEMI and MVO ( 14 16 ). However, IMR is typically measured using a thermodilution wire with a sensor near its tip, which has less torquability in comparison with workhorse wires, leading to technical difficulty and increased cost ( 17 ). Furthermore, IMR requires pharmacological induction of hyperemia, which may be uncomfortable for patients and has a risk of morbidity from arrhythmia.…”
Section: Discussionmentioning
confidence: 99%
“…These systems have been well validated, with a reduction in pressure drift reported with the optical method[ 30 , 31 ]. Moreover, the latest versions of optical pressure guidewires enable its use as a workhorse guidewire for PCI by disconnecting and reconnecting the system.…”
Section: Fractional Flow Reservementioning
confidence: 99%
“…Several intracoronary pressure systems are currently available on the market: WaveWire™ (Philips Inc, Eindhoven, the Netherlands), COMET™ (Boston Scientific, Marlborough, MA, USA), PressureWire™ (St. Jude/Abbott, St. Paul, MN, United States), NAVVUS RXi ® (ACIST Medical Systems, Inc, Eden Prairie, MN), TruePhysio™ (Insight Lifetech) and OptoWire Deux (Opsens medical, Quebec, Canada) with fewer drift[ 31 ]. The COMET trial comparing the COMET™ and the PressureWire™ found similar results in terms of safety and performance[ 34 ].…”
Section: Fractional Flow Reservementioning
confidence: 99%
“…However, if the gradient exceeds 2-3 mmHg, it is considered a significant drift 3,5 . In this case, repeated pressure equalization and function testing should be performed to obtain an accurate measurement.The prevalence of significant drift ranges from 7.4 to 73% for piezoelectric sensors 4,[6][7][8][9] . This large variation may be due to the differences in vendor pressure wires, the definitions of significant drifts, and operator-dependent errors.…”
mentioning
confidence: 99%
“…The prevalence of significant drift ranges from 7.4 to 73% for piezoelectric sensors 4,[6][7][8][9] . This large variation may be due to the differences in vendor pressure wires, the definitions of significant drifts, and operator-dependent errors.…”
mentioning