2018
DOI: 10.1159/000492674
|View full text |Cite
|
Sign up to set email alerts
|

Feasibility of International Proposed Standardized Enteral Connector for Semi-Solid Formula Feeding

Abstract: Background/Aims: The current study was undertaken to assess if the semi-solid formulas could be used with a new ENFit connector with similar force to current percutaneous endoscopic gastrostomy (PEG) tubes. Methods: Experiment 1: We measured the applied pressure (force) needed to compress the syringe containing 7 viscous semi-solid formulas with a 20 Fr PEG tube and low-profile tube through the ENFit connector or the current connector. Experiment 2: This experiment was conducted to evaluate the compression for… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2022
2022
2022
2022

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(4 citation statements)
references
References 15 publications
(25 reference statements)
0
4
0
Order By: Relevance
“…Concerns have been raised by some stakeholders about the performance of the ENFit‐based G‐tube devices in delivering HEN 6 . HEN has been increasingly used because it reduces the time spent in a hospital 10 . HEN users have raised concerns that the new G‐tube devices will result in increased time of feeding, increased likelihood of clogging, difficulty in pushing diets because of a higher resistance to flow, cleaning problems, and the potential for drug dosing inaccuracy because of ENFit's smaller inner diameter and direction of flow 4,13 .…”
Section: The Changing Landscape Of Home Enteral Nutrition (Hen)mentioning
confidence: 99%
See 3 more Smart Citations
“…Concerns have been raised by some stakeholders about the performance of the ENFit‐based G‐tube devices in delivering HEN 6 . HEN has been increasingly used because it reduces the time spent in a hospital 10 . HEN users have raised concerns that the new G‐tube devices will result in increased time of feeding, increased likelihood of clogging, difficulty in pushing diets because of a higher resistance to flow, cleaning problems, and the potential for drug dosing inaccuracy because of ENFit's smaller inner diameter and direction of flow 4,13 .…”
Section: The Changing Landscape Of Home Enteral Nutrition (Hen)mentioning
confidence: 99%
“…Low‐profile G‐tube devices are also considered to have a lower likelihood of dislodgement compared with standard G‐tubes. The Japanese Pharmaceuticals and Medical Device Agency also conducted a similar force study, 10 although their testing was performed on ENFit prototypes rather than finished devices. They used seven different commercial semisolid formulas and studied push rates of 0.83, 1.66, and 2.5 ml/s.…”
Section: Review Of Literaturementioning
confidence: 99%
See 2 more Smart Citations