Fast x-ray detector system with simultaneous measurement of timing and energy for a single photon

Abstract: We developed a fast X-ray detector system for nuclear resonant scattering (NRS) experiments. Our system employs silicon avalanche photo-diode (Si-APD) as a fast X-ray sensor. The system is able to acquire both timing and energy of a single X-ray photon simultaneously in a high rate condition, 10 counts per second for one Si-APD. The performance of the system was investigated in SPring-8, a synchrotron radiation facility in Japan. Good time resolution of 120 ps (FWHM) was achieved with a slight tail distributio… Show more

Help me understand this report
View preprint versions

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
10
0

Year Published

2017
2017
2024
2024

Publication Types

Select...
5
2
1

Relationship

2
6

Authors

Journals

citations
Cited by 17 publications
(10 citation statements)
references
References 31 publications
0
10
0
Order By: Relevance
“…The conceptual diagram of the signal processing is shown in Fig 6 . Each Si-APD output is amplified by a fast amplifier (Mini-circuits, RAM-8A+) located directly behind the Si-APDs. The amplified output is processed in several ways to obtain event characteristics 19,20 . First of all, it is converted to a logic signal by a lab-built constant fraction discriminator (CFD); this yields an event occurrence time.…”
Section: Appendices a Detector System A1 Apd Detector And Readout Systemmentioning
confidence: 99%
“…The conceptual diagram of the signal processing is shown in Fig 6 . Each Si-APD output is amplified by a fast amplifier (Mini-circuits, RAM-8A+) located directly behind the Si-APDs. The amplified output is processed in several ways to obtain event characteristics 19,20 . First of all, it is converted to a logic signal by a lab-built constant fraction discriminator (CFD); this yields an event occurrence time.…”
Section: Appendices a Detector System A1 Apd Detector And Readout Systemmentioning
confidence: 99%
“…The energy spectra obtained in the actual NRS measurement at high count rate of more than 10 6 cps have been reported in our previous paper [15]. This is potentially useful for other types of measurements as well as the 229 Th NRS measurement, for example, depth selective Mössbauer spectroscopy by electron detection proposed in [24].…”
Section: Energy Measurement Methodsmentioning
confidence: 52%
“…We recently developed a fast single X-ray photon detection system by using thin Si-APDs [15]. It was developed for the synchrotron-radiation based nuclear resonant scattering (NRS) measurement of thorium-229 [16].…”
Section: Introductionmentioning
confidence: 99%
“…The pellet was covered with two 20 Â 20 mm MgF 2 substrates, one on either side. The scattered X-ray photons were detected by a dedicated energy-sensitive X-ray detector (Masuda et al, 2017(Masuda et al, , 2019a.…”
Section: Nrs Measurementmentioning
confidence: 99%