1983
DOI: 10.1177/106002808301700710
|View full text |Cite
|
Sign up to set email alerts
|

Falsely Elevated Aminoglycoside Serum Levels in Jaundiced Patients

Abstract: Two cases in which hyperbilirubinemia produced falsely elevated aminoglycoside concentrations, using the Emit assay, are described. Patients' sera were analyzed by both the Emit and a microbiologic assay. The first patient had a 26-percent greater tobramycin concentration with the Emit assay, as compared with the microbiologic assay, when the total bilirubin was 20.1 mg%. For the same patient, there was essentially no difference between the two assays when the total bilirubin was 4 mg%. The Emit concentrations… Show more