1992
DOI: 10.1016/0168-9002(92)90135-q
|View full text |Cite
|
Sign up to set email alerts
|

Experimental evaluation of TDCR models for the 3 PM liquid scintillation counter

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
12
0

Year Published

1994
1994
2024
2024

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 64 publications
(13 citation statements)
references
References 3 publications
0
12
0
Order By: Relevance
“…In the Hidex 300SL LSA, a standard quench curve can be stored with the curve fit option, which can be automatically employed to determine the real counting efficiency based on the TDCR value of the sample. The TDCR efficiency calculation technique has been approved to be suitable for routine analysis of beta-emitting nuclides in samples with different quench levels [30][31][32]. The TDCR based LSC method has been reported in the 1980's [33], since then many laboratories have built their own TDCR system for LSC measurement.…”
Section: Tdcr Liquid Scintillation Countingmentioning
confidence: 99%
See 2 more Smart Citations
“…In the Hidex 300SL LSA, a standard quench curve can be stored with the curve fit option, which can be automatically employed to determine the real counting efficiency based on the TDCR value of the sample. The TDCR efficiency calculation technique has been approved to be suitable for routine analysis of beta-emitting nuclides in samples with different quench levels [30][31][32]. The TDCR based LSC method has been reported in the 1980's [33], since then many laboratories have built their own TDCR system for LSC measurement.…”
Section: Tdcr Liquid Scintillation Countingmentioning
confidence: 99%
“…The TDCR based LSC method has been reported in the 1980's [33], since then many laboratories have built their own TDCR system for LSC measurement. However, these early applications of the TDCR method was exclusively employed in metrological laboratories for radionuclide standardization [31,32,[34][35][36][37]. The first commercial TDCR Lead shield 3 PMTs…”
Section: Tdcr Liquid Scintillation Countingmentioning
confidence: 99%
See 1 more Smart Citation
“…All ampoules were checked for total activity in ionization chambers and radiochemical impurities were determined by y-ray spectrometry. 241 Am was measured by liquid scintillation (TDCR method [11]). …”
Section: Radionuclides Solutions and Chemicalsmentioning
confidence: 99%
“…The TDCR LSC utilizes 3 symmetrically arranged photomultiplier tubes (PMT) to get the triple coincidence counts and 3 double coincidence counts. In 1992, the TDCR with its counting efficiency calculated was first reported by LNHB in France [1] and LRS in Poland [2] . Subsequently, more than ten national metrology labs, such as NIST [3] , ANSTO [4] , PTB [5] , have adopted TDCR LSC as the primary standardization for pure β nuclide activity measurement.…”
Section: Introductionmentioning
confidence: 96%