Evaluations of UltraiQ software for objective ultrasound image quality assessment using images from a commercial scanner

Abstract: We evaluated a commercially available software package that uses B‐mode images to semi‐automatically measure quantitative metrics of ultrasound image quality, such as contrast response, depth of penetration (DOP), and spatial resolution (lateral, axial, and elevational). Since measurement of elevational resolution is not a part of the software package, we achieved it by acquiring phantom images with transducers tilted at 45 degrees relative to the phantom. Each measurement was assessed in terms of measurement … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
4
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
6

Relationship

2
4

Authors

Journals

citations
Cited by 7 publications
(4 citation statements)
references
References 12 publications
0
4
0
Order By: Relevance
“…An experimental investigation of scanner’s resolution in three space directions for 3 MHz phased array transducers was presented in Long et al [ 51 ]. The individual measurements as a function of depth agree with expectation: elevational resolution is worst (minimum 2 mm) and varies the most versus depth, with a distinct hourglass profile; the axial resolution is best and remains roughly constant (around 1 mm) versus depth; the lateral resolution is between the other two and demonstrates modest changes versus depth.…”
Section: Discussionmentioning
confidence: 99%
“…An experimental investigation of scanner’s resolution in three space directions for 3 MHz phased array transducers was presented in Long et al [ 51 ]. The individual measurements as a function of depth agree with expectation: elevational resolution is worst (minimum 2 mm) and varies the most versus depth, with a distinct hourglass profile; the axial resolution is best and remains roughly constant (around 1 mm) versus depth; the lateral resolution is between the other two and demonstrates modest changes versus depth.…”
Section: Discussionmentioning
confidence: 99%
“…For each device, six B-mode images of the phantom were taken using the abdominal preset and a curved array or a curved preset on a linear transducer (for the Butterfly device). This approach was reported to produce high inter- and intraoperator repeatability 16 . Full width at half maximum (FWHM) of the embedded pins was calculated automatically using the UltraiQ Desktop software.…”
Section: Methodsmentioning
confidence: 99%
“…Gray map function was plotted using a custom MATLAB program in order to identify a map function on system 1 which would be close to system 2. A CIRS Model 040GSE phantom (CIRS Inc., Norfolk, VA) was used to measure global contrast during the parameter changes with the gray‐scale targets (−9, −6, −3, +3, +6 dB) using the UltraiQ software (Cablon Medical B.V., Leusden, The Netherlands), 7 in order to confirm that preset changes were moving toward the right direction, that is, improving contrast as identified in step 1. If there is no reference system to try to move toward, radiologists and sonographers might need to be present or provide timely input, especially for subjective preferences such as tint map. Fine‐tuning basic parameters and more advanced parameters such as spatial compounding, speckle reduction, and noise reduction, using volunteers and/or good anthropomorphic phantoms if volunteer scanning is not feasible.…”
Section: Methodsmentioning
confidence: 99%