2020
DOI: 10.1515/cclm-2020-0848
|View full text |Cite
|
Sign up to set email alerts
|

Evaluation of the first immunosuppressive drug assay available on a fully automated LC-MS/MS-based clinical analyzer suggests a new era in laboratory medicine

Abstract: ObjectivesDue to its high specificity, liquid chromatography-tandem mass spectrometry (LC-MS/MS) is considered the gold standard in diagnostic areas such as therapeutic monitoring of immunosuppressive drugs (ISDs). However, many laboratories still rely on immunoassays for ISD quantification in a tradeoff between analytical performance and the advantages of fully automated analyzers – shorter turnaround times, greater ease of use, and 24/7 availability.MethodsThe LC-MS/MS-based Thermo Scientific™ Cascadion™ SM … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

1
10
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
6
2
1

Relationship

1
8

Authors

Journals

citations
Cited by 27 publications
(12 citation statements)
references
References 17 publications
1
10
0
Order By: Relevance
“…In a smaller study that included only 21 patient samples, the observed bias against the LC-MS/MS system was not detectable. 17 This was attributed to the low number of samples tested. Having a bias of −9% in the case of CSA could result in a certain number of patient results not being detected as above therapeutic range, and thus, dose correction may be missed out for those patients.…”
Section: Discussionmentioning
confidence: 99%
“…In a smaller study that included only 21 patient samples, the observed bias against the LC-MS/MS system was not detectable. 17 This was attributed to the low number of samples tested. Having a bias of −9% in the case of CSA could result in a certain number of patient results not being detected as above therapeutic range, and thus, dose correction may be missed out for those patients.…”
Section: Discussionmentioning
confidence: 99%
“…Thirdly, the metabolites described here can be analyzed in a single LC-MS run. Recently a fully automatic LC-MS system, suitable for 24/7 use, was introduced into routine diagnostics ( 74 ). This instrument, which is easy to operate and does not call for specific LC-MS expertise ( 74 ), not only opens up perspectives for every routine diagnostic laboratory to apply LC-MS, but also analyzes complex diagnostic patterns in a fully automatic, high-throughput manner in the future.…”
Section: Discussionmentioning
confidence: 99%
“…Examples of automated systems that are currently available are the Shimadzu CLAM System 145 and ThermoFisher Scientific's Cascadion™ SM Clinical Analyzer 146 . The latter has been tested and compared with routine tests for Vitamin D and immunosuppressive drugs (cyclosporine A, sirolimus, tacrolimus, and everolimus) and showed excellent analytical performance 147,148 . For Vitamin D, all samples were within 10% of the reference concentrations, and for the immunosuppressive drugs, the mean deviation from the immunoassays was 17%–19%, 147,148 suggesting that automated LC–MS provides robust clinical data.…”
Section: Techniques and Challenges For Tdmmentioning