Evaluation of pull‐off strength and seating displacement of sleeved ceramic revision heads in modular hip arthroplasty

Abstract: Corrosion in revision total hip arthroplasty can be mitigated using a ceramic head on a well‐fixed in situ stem, but concerns of their early failure because of any surface defects on in situ stem necessitates the use of a titanium sleeve, which furnishes a factory‐finish surface. These sleeves are manufactured in different sizes allowing neck‐length adjustment. The strength of the taper junction of non‐sleeved primary heads is well‐investigated, but the influence of an interposed titanium sleeve on achieving a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(2 citation statements)
references
References 11 publications
0
2
0
Order By: Relevance
“…Furthermore, some studies describe a force of 4 kN as the optimum impaction force for head taper [10,28,29,31,32], although the ISO 7206-10 [25] recommends a force of 2 kN. Various conditions influence the forces applied to the joint connection, such as the stiffness of the impact tip of the instrument [29] or the implant component size [33], so further investigations and detailed dimensioning are required to define the optimum joining force. Compensation could be achieved by using different tips or an adjustable force range of the instrument.…”
Section: Discussionmentioning
confidence: 99%
“…Furthermore, some studies describe a force of 4 kN as the optimum impaction force for head taper [10,28,29,31,32], although the ISO 7206-10 [25] recommends a force of 2 kN. Various conditions influence the forces applied to the joint connection, such as the stiffness of the impact tip of the instrument [29] or the implant component size [33], so further investigations and detailed dimensioning are required to define the optimum joining force. Compensation could be achieved by using different tips or an adjustable force range of the instrument.…”
Section: Discussionmentioning
confidence: 99%
“…When coupling the ceramic head to the existing trunnion, a titanium adapter sleeve is recommended to optimize taper fit and decrease material transfer because the sleeve provides a factory-finish surface in case of any surface defects on the in situ stem. 31,32 The one exception to this guideline occurs when treating ALTR in CoCr stem/CoCr head couples. In such cases, the ceramic head should be directly impacted onto the trunnion without an adapter sleeve because the titanium sleeve would introduce a mixed-metal couple, which increases the risk of metallosis.…”
Section: Mechanically Assisted Crevice Corrosionmentioning
confidence: 99%