2022
DOI: 10.1186/s13014-021-01970-8
|View full text |Cite
|
Sign up to set email alerts
|

Evaluation of plan robustness on the dosimetry of volumetric arc radiotherapy (VMAT) with set-up uncertainty in Nasopharyngeal carcinoma (NPC) radiotherapy

Abstract: Purpose To evaluate the sensitivity to set up the uncertainty of VMAT plans in Nasopharyngeal carcinoma (NPC) treatment by proposing a plan robustness evaluation method. Methods 10 patients were selected for this study. A 2-arc volumetric-modulated arc therapy (VMAT) plan was generated for each patient using Varian Eclipse (13.6 Version) treatment planning system (TPS). 5 uncertainty plans (U-plans) were recalculated based on the first 5 times set-… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
3
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
4
1
1

Relationship

0
6

Authors

Journals

citations
Cited by 7 publications
(3 citation statements)
references
References 34 publications
0
3
0
Order By: Relevance
“…The average EUD values for femoral head-R in the VMAT plans without error 10.29 Gy were higher than those in the VMAT plans with error 10.61 Gy the average NTCP values of femoral head-R without error in the VMAT plans, 0.0% and with error in the VMAT plans 0.0% are shown in Table (17). For dosimetric response, The average volume in VMAT plans without errors is 171.2 cc, the average minimum dose is 2.20 Gy, the average maximum dose is 35.01 Gy, the average mean dose is 16.65 Gy, the average modal dose is 13.07 GY, the average of median dose 17.59 Gy, the average of STD 7.04 Gy, and the average of equivalent sphere diameter 6.9cm are shown in Table (18).…”
Section: Right Femoral Headmentioning
confidence: 85%
See 2 more Smart Citations
“…The average EUD values for femoral head-R in the VMAT plans without error 10.29 Gy were higher than those in the VMAT plans with error 10.61 Gy the average NTCP values of femoral head-R without error in the VMAT plans, 0.0% and with error in the VMAT plans 0.0% are shown in Table (17). For dosimetric response, The average volume in VMAT plans without errors is 171.2 cc, the average minimum dose is 2.20 Gy, the average maximum dose is 35.01 Gy, the average mean dose is 16.65 Gy, the average modal dose is 13.07 GY, the average of median dose 17.59 Gy, the average of STD 7.04 Gy, and the average of equivalent sphere diameter 6.9cm are shown in Table (18).…”
Section: Right Femoral Headmentioning
confidence: 85%
“…In literature, some works were focused on the dosimetric or radiobiological assessment of systematic isocenter shift errors for different types of cancers. [17,18,24]. This work aimed to evaluate both the dosimetric and radiobiological effects of the random setup error that usually occurs during the position of prostate patients.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation