Evaluation of nanoDot optically stimulated luminescence dosimeter for cone-shaped small-field dosimetry of cyberknife stereotactic radiosurgery unit: A monte carlo simulation and dosimetric verification study

Abstract: Aim: The aim of this study was to investigate the adequacy of nanoDot optically stimulated luminescence (OSL) dosimeter for small field dosimetry before its in vivo applications in CyberKnife SRS unit. Materials and Methods: A PTW 60018 SRS Diode, 60019 microDiamond, and Gafchromic EBT3 films were used along with a nanoDot carbon-doped aluminum oxide OSL dosimeter to collect and compare beam data. In addition, the EGSnrc/BEAMnrc code was emplo… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
3
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
4
1
1

Relationship

1
5

Authors

Journals

citations
Cited by 6 publications
(4 citation statements)
references
References 43 publications
0
3
0
Order By: Relevance
“…The water-equivalent window thickness of the microdiamond is smaller; however, the difference between their correction factors we obtained for another study was not significant. 14 In this work, measured values for field sizes were the same except for the 15 mm cone. For the smallest cone, there is a direction of difference between these two evaluations.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…The water-equivalent window thickness of the microdiamond is smaller; however, the difference between their correction factors we obtained for another study was not significant. 14 In this work, measured values for field sizes were the same except for the 15 mm cone. For the smallest cone, there is a direction of difference between these two evaluations.…”
Section: Discussionmentioning
confidence: 99%
“…The MC simulation was validated through percentage depth dose (PDD) and output factors measurements and results were published elsewhere. 14 For MC dose profiles the water phantom was modelled as 30 cm × 30 cm × 30 cm using different voxel sizes, 1 mm for all cone sizes except 60 mm cone for which it was 5 mm. The dose profile computation was at the depth of 5 cm with SAD of 80 cm.…”
Section: Egsnrc Monte Carlo Simulationsmentioning
confidence: 99%
“…It boasts a lower detection limit of 0.1 mGy and an accuracy of 10% in energy dependence within the diagnostic energy range of 70 kVp to 140 kVp, and standard accuracy within 10%. In addition, experimental data have revealed 40-50 times greater sensitivity for OSLDs compared with TLDs [20][21][22].…”
Section: Methodsmentioning
confidence: 99%
“…Currently, OSLD is commercially available by Nagase Landauer, Tsukuba, Ibaraki, Japan, and widely used in various studies, including environmental, personnel dosimeter, and patients dose measurements in both therapeutic and diagnostic radiology (17)(18)(19).…”
Section: Technology Of Osldmentioning
confidence: 99%