Evaluation of four-terminal-pair capacitance standards using electrical equivalent circuit model

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
4
0

Year Published

2015
2015
2022
2022

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 15 publications
(5 citation statements)
references
References 13 publications
0
4
0
Order By: Relevance
“…The frequency characteristics of the 4TP capacitance standard up to 10 MHz are also estimated by determining capacitance deviation from the reference capacitance to the typical capacitance values determined at frequencies above 1 kHz. To validate the automated characterization module, the values of capacitance standards (100 pF and 1000 pF) up to 10 MHz computed through the automated characterization module are compared with the evaluated values obtained by s-parameter method [11] and also with the values obtained by measuring each 4TP standard using an Agilent 4285A LCR meter. The Table illustrates the comparison results for 1000 pF and 100 pF capacitance standards along with associated expanded uncertainty in measurement (u e ) up to 10 MHz.…”
Section: Characterization Resultsmentioning
confidence: 99%
See 2 more Smart Citations
“…The frequency characteristics of the 4TP capacitance standard up to 10 MHz are also estimated by determining capacitance deviation from the reference capacitance to the typical capacitance values determined at frequencies above 1 kHz. To validate the automated characterization module, the values of capacitance standards (100 pF and 1000 pF) up to 10 MHz computed through the automated characterization module are compared with the evaluated values obtained by s-parameter method [11] and also with the values obtained by measuring each 4TP standard using an Agilent 4285A LCR meter. The Table illustrates the comparison results for 1000 pF and 100 pF capacitance standards along with associated expanded uncertainty in measurement (u e ) up to 10 MHz.…”
Section: Characterization Resultsmentioning
confidence: 99%
“…Capacitance from Computed s-parameter capacitance using Measured capacitance method [11] automated module using LCR meter 1000 pF 100 pF 1000 pF 100 pF 1000 pF ± U e 100 pF ± U e 1other. The automated module will be further modified to incorporate the extended feature of the evaluation of expanded uncertainty in the measurement of each capacitance standard.…”
Section: Freq Mhzmentioning
confidence: 99%
See 1 more Smart Citation
“…The LCR meter supports 4TP measurement, which minimises the prevalent secondary effects due to stray capacitances and mutual or self-inductances of connecting leads. In addition, it is metrologically traceable to the National Standard of Impedance maintained at the "National Measurement Institute" of India, the CSIR-National Physical Laboratory (CSIR-NPL) [23][24][25][26]. The LDTF is fitted with 4TP Sub-Miniature versions A (SMA) connectors to support 4TP measurement as shown in Figure 1.…”
Section: Experimental Set-upmentioning
confidence: 99%
“…The user must select the proper equivalent circuit mode to minimize measurement error. Figure 9 shows the description of the four terminal pair measurement circuit used for the data acquisition process [62,63].…”