Abstract:The objective of this work is to check the dosimetric performances of the TLD-100 as stated by the manufacturer as well as the technical standards of radiation protection. The purpose of the performance audit is to assess the inhomogeneity of TLD sensitivity, repeatability and reproducibility, linearity, energy dependence, angular dependence, and fading. All tests were performed under the conditions of ambient temperature and relative humidity recommended by the manufacturer. We began the study by calibrating … Show more
“…Sabar et al, likewise, reported that there could be very high deviation (above 40%) below 0.3mGy (18). This was also supported by Kouakou et al, who evaluated dosimetric performance and uncertainty for TLD-100, and reported that a proportion of TL chips were unable to measure accurately at 0.1-10mGy (19). A study by Kadir et al investigated the uncertainty associated with the energy response of TLD-100 to an ideal dose, which had a maximum deviation of 125.04% for Hp (0.07) against 45.22% for Hp (10) at 24keV, with the shallow dose exhibiting the higher uncertainty (20).…”
Purpose: The purpose of this study was to compare calibration factors for deep dose equivalent Hp (10) and shallow dose equivalent Hp (0.07) between Cesium (Cs)-137 and X-ray sources when they are exposed to same dose and to determine uncertainties with MTS-N (LiF: Mg, Ti) chips when they are exposed to low dose ≤ 2mGy.
Material and Methods: Thermoluminescent (TL) chips were annealed at 400oC for one hour and allowed to cool and were subjected to a temperature of 100oC for another two hours using a TLD Furnace Type LAB-01/400. They were then taken to a Secondary Standard Dosimetry Laboratory (SSDL) for irradiation using a Cs-137 source at known doses (0.2-2mGy). A RadPro Cube 400 manual TLD Reader was used to determine corresponding TL signal. The above process was replicated but with a calibrated X-ray unit as the source for calibration.
Results: The calibration factors (CF) from the line graph of dose (mGy) against TL signal (count) for Cs-137 source with Hp (10) and Hp (0.07) were 3.72 x 10-6 and 5.97x10-6 mGy/count respectively. Those with X-ray source for Hp (10) and Hp (0.07) were 3.44x10-6 and 4.05x10-6 mGy/count respectively with an overall coefficient of determination (R2) = 0.99. The adjusted maximum percentage deviation between the actual and calculated dose for both sources was -2.74%. The percent (%) deviation of the mean with both sources for Hp (10) and Hp (0.07) was 3.9% and 19% respectively.
Conclusion: Adjusted percent deviation from both sources were within the recommended dose limit of ±30% by the Radiological Protection Institute of Ireland (RPII) and within the International Commission on Radiological Protection (ICRP) limit respectively. Better accuracy was seen for Hp (10) with both sources compared to Hp (0.07). Calibration of the MTS-N chips using both sources was successful and can be used for personal dosimetry.
“…Sabar et al, likewise, reported that there could be very high deviation (above 40%) below 0.3mGy (18). This was also supported by Kouakou et al, who evaluated dosimetric performance and uncertainty for TLD-100, and reported that a proportion of TL chips were unable to measure accurately at 0.1-10mGy (19). A study by Kadir et al investigated the uncertainty associated with the energy response of TLD-100 to an ideal dose, which had a maximum deviation of 125.04% for Hp (0.07) against 45.22% for Hp (10) at 24keV, with the shallow dose exhibiting the higher uncertainty (20).…”
Purpose: The purpose of this study was to compare calibration factors for deep dose equivalent Hp (10) and shallow dose equivalent Hp (0.07) between Cesium (Cs)-137 and X-ray sources when they are exposed to same dose and to determine uncertainties with MTS-N (LiF: Mg, Ti) chips when they are exposed to low dose ≤ 2mGy.
Material and Methods: Thermoluminescent (TL) chips were annealed at 400oC for one hour and allowed to cool and were subjected to a temperature of 100oC for another two hours using a TLD Furnace Type LAB-01/400. They were then taken to a Secondary Standard Dosimetry Laboratory (SSDL) for irradiation using a Cs-137 source at known doses (0.2-2mGy). A RadPro Cube 400 manual TLD Reader was used to determine corresponding TL signal. The above process was replicated but with a calibrated X-ray unit as the source for calibration.
Results: The calibration factors (CF) from the line graph of dose (mGy) against TL signal (count) for Cs-137 source with Hp (10) and Hp (0.07) were 3.72 x 10-6 and 5.97x10-6 mGy/count respectively. Those with X-ray source for Hp (10) and Hp (0.07) were 3.44x10-6 and 4.05x10-6 mGy/count respectively with an overall coefficient of determination (R2) = 0.99. The adjusted maximum percentage deviation between the actual and calculated dose for both sources was -2.74%. The percent (%) deviation of the mean with both sources for Hp (10) and Hp (0.07) was 3.9% and 19% respectively.
Conclusion: Adjusted percent deviation from both sources were within the recommended dose limit of ±30% by the Radiological Protection Institute of Ireland (RPII) and within the International Commission on Radiological Protection (ICRP) limit respectively. Better accuracy was seen for Hp (10) with both sources compared to Hp (0.07). Calibration of the MTS-N chips using both sources was successful and can be used for personal dosimetry.
“…The Harshaw 6600 Plus system, consisting of a reader and dosimeters whose performance characteristics and overall uncertainty [4] [5] were evaluated before use, allowed the doses received by the operators to be assessed according to their position in the room. For the dosimetric evaluation, each operator was equipped with one or more dosimeters to assess the effective dose and equivalent doses to the extremities and lens of the eye for the radiological procedure performed.…”
Ionizing radiation is widely used in medical practice for both diagnostic and therapeutic purposes. However, they are not devoid of adverse health effects on exposed operators. Therefore, we propose to assess the radiological risk of the interventional radiology team of the International Polyclinic of Indénié of Abidjan during the procedures of Embolization of uterine fibromyomas. The effective and equivalent doses to the extremities and crystals were estimated by TLD dosimeters positioned at different body areas selected on the basis of a radiological analysis of the exposure of personnel during five procedures. The analysis of fluoroscopic and record times showed variability in or an average fluoroscopy time of 32.37 min with extremes of 25.14 to 56.32 min; average record time of 0.52 min with extremes of 0.12 min to 1 min. The annual effective doses were respectively, 4.04 mSv, 3.42 mSv, 2.84 mSv; 2.28 mSv, in the Radiologist (R), Operator Assistant (O2); Anesthesiologist (A2), Manipulator (M2). The annual equivalent doses to the radiologist's predominant extremities and lenses were left index and left lens with values of 37.07 mSv and 9.46 mSv. The estimated doses in our study are reassuring from a regulatory point of view of dose limits. The results of our work have shown no significant short-term danger to the health of personnel.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.